Article count:1333 Read by:1940015

Account Entry

【ShiShuo Design】Come and see how to design the best front end for UHF partial discharge online monitoring system~

Latest update time:2021-11-24 10:24
    Reads:




According to IEC 60270, partial discharge (PD) is an electrical discharge that occurs in a partially insulated area between two conductive electrodes with a gap. Partial discharge is widely considered to be the best early warning indicator of insulation degradation in electrical assets within a power grid.




When partial discharge occurs, a signal with a wide frequency range is generated, so there are 4 partial discharge detection technologies for different frequency ranges. Ultrasonic detection technology is for the frequency range of 20 kHz to ~200 kHz, high-frequency current transformer (HFCT) detection technology is for the frequency range of 3 MHz to ~30 MHz, transient earth voltage (TEV) detection technology is for the frequency range of 3 MHz to ~100 MHz, and ultra-high frequency (UHF) detection technology is for the frequency range of 300 MHz to ~1500 MHz. UHF detection technology has high detection sensitivity and is widely used in partial discharge online monitoring systems for gas insulated switchgear (GIS), transformers and ring main units (RMUs).


Partial discharge signal analysis

According to Section 7.1 of the Q/GDW11282-2014 standard "Specifications for Field Detection of Partial Discharge UHF Couplers for Gas-Insulated Metal-Enclosed Switchgear", a standard PD signal generator can generate the following PD pulse signal characteristics: the pulse rise time does not exceed 300 ps, ​​and the pulse width is between 10 ns and 500 ns. This information is then used to build a PD simulator signal in Python. The rise time is 300 ps and the fall time is 10 ns. The peak amplitude of the pulse signal is 100 mV and the peak-to-peak noise is 10 mV. The sampling rate is 10 GSPS and the sampling time is 10 μs. The pulse is placed in the middle of the sampling time, and both the rising and falling waveforms are linearly fitted.


The simulated PD signal time domain waveform is shown in Figure 1, and the frequency domain waveform is shown in Figure 2. According to Figure 2, the PD signal with the highest energy is in the frequency range below 1 GHz. When the pulse rise time is less than 300 ps, ​​more energy is distributed in the higher frequency range.


Figure 1. PD signal time domain waveform.


Figure 2. PD signal frequency domain waveform.


In the modern complex electromagnetic environment, there are many wireless interference signals between UHF PDs with operating frequencies between 300 MHz and 1500 MHz. In order to eliminate this interference, customers generally choose sub-bands between 300 MHz and 1.5 GHz to capture PD pulses. Under normal circumstances, the wireless communication signal of GSM around 900 MHz will be the largest interference signal. One way to solve this problem is to use a band-stop filter (BRF) to suppress signals from 800 MHz to 1000 MHz. The typical sub-band division scheme is shown in Table 1. Of course, the sub-band division is flexible, and customers can adjust it according to the actual electromagnetic environment.


Table 1. Typical UHF PD sub-band division scheme


According to the sub-band division in Table 1, we only retain the corresponding energy spectrum components of the PD signal spectrum shown in Figure 2, and then perform an inverse fast Fourier transform (IFFT) to study what the time domain waveform will look like after the corresponding filtering. The filtered time domain waveform is shown in Figure 3. According to Figure 3, after filtering, the PD pulse peak value will decrease. After filtering, the PD pulse rise time will increase and the fall time will decrease. After filtering, among all waveforms, the full frequency band has the largest peak, followed by the band stop band and the low pass band. The high pass band has the smallest peak, but the PD pulse can still be captured.


Figure 3. Time domain waveform of the PD signal after filtering.


UHF PD Detection RF Front End Using ADI Signal Chain

A UHF PD detection RF front-end board with 4 channels can be developed using the ADI signal chain. The block diagram of one channel is shown in Figure 4, and the front view of the entire board is shown in Figure 5.

Figure 4. UHF PD detection RF front-end board block diagram.


Figure 5. Front view of the UHF PD detection RF front-end board.


The first stage of this front end is the RF gain block ADL5611. The ADL5611 has a low noise figure (NF) of 2.1 dB and a high P1dB of 21 dBm, which provides a high dynamic range. The ADL5611 has a gain of 22 dB and its gain is extremely flat within the UHF PD operating frequency of 300 MHz to 1500 MHz, with a gain ripple of less than 0.4 dB. All these features make the ADL5611 very suitable for UHF PD detection applications.


The second stage is an LC-based 300 MHz to 1500 MHz bandpass filter (BPF) that provides out-of-band interference rejection.


The third stage uses two single-pole four-throw (SP4T) RF switches HMC7992 to implement the band selection circuit. The first RF path is a DC to 800 MHz low-pass path, the second RF path is a 1 GHz high-pass path, the third path is a band-stop path from 800 MHz to 1 GHz, and the fourth path is a straight-through path. Depending on the different RF path selections, customers can choose different RF bands to capture PD pulses in a band with no interference or minimal interference. The HMC7992 has a low insertion loss of 0.6 dB, a high isolation of 45 dB, and a high P0.1dB of 33 dBm.


Stage 4 is a 300 MHz to 1500 MHz BPF, the same BPF used in stage 2, which further provides out-of-band interference suppression.


The last stage is the RF logarithmic detector ADL5513, which converts the UHF PD signal into a low-frequency signal of tens of MHz. Therefore, an ADC with a sampling rate of 40 MSPS or 65 MSPS can be used to convert the analog PD signal into a digital signal. For PD detection applications, the main characteristics of the RF detector required are response time and dynamic range. The ADL5513 has a response time as low as 20 ns and a dynamic range as high as 80 dB, so it is very suitable for PD detection applications. The RF logarithmic detector AD8318 is also suitable for PD detection applications. Compared with the ADL5513, it has a faster response time but a slightly smaller dynamic range.


Test Results

The key performance of the board was tested and Figures 6 to 8 are screen shots.


Figure 6 shows the S parameters of the through path from the first-stage input to the input port of the last-stage ADL5513. The figure shows that the gain is about 14 dB, the gain flatness is better than 2 dB, and the input return loss is better than –8 dB over the full frequency band from 300 MHz to 1500 MHz.

Figure 6. From the first stage input to the final stage ADL5513 input.

S parameters in the through path.


Figure 7 shows the response curve of the output voltage to the power of the input CW signal at the center frequency of the PD at 900MHz. Two channels were measured using input power. According to the test results, the entire signal chain has a linear response in the input power range of –75 dBm to –5 dBm. The performance consistency between channels is also very good.


Figure 7. Output voltage vs. input power.


Figure 8 shows the output waveform measured when a 900 MHz continuous wave signal pulse is input. The signal power is –75 dBm, the pulse width is 5 μs, and the pulse period is 10 μs. According to the waveform, when the signal power is as low as –75 dBm, the signal-to-noise ratio of the output signal is still considerable.

Figure 8. Output response to a –75 dBm pulsed CW input.


in conclusion

This article shows how to use the ADI signal chain to build a UHF PD detection board. This complete reference design allows users to flexibly select different frequency bands to eliminate interference in complex electromagnetic environments. It also meets the requirements of the Chinese Q/GDW11059.8-2013 standard.





Original article from Analog Devices


Scan the QR code to purchase immediately ↑ ↑ ↑



About Shijian

Leading authorized distributor of components in Asia Pacific


Excelpoint is a complete solutions provider, providing high-quality components, engineering design and supply chain management services to Asian electronics manufacturers including original equipment manufacturers (OEM), original design manufacturers (ODM) and electronic manufacturing service providers (EMS).


World Health is a listed company on the main board of Singapore with a history of more than 30 years. World Health China is headquartered in Hong Kong and currently has more than ten branches and offices in China, covering major large and medium-sized cities in China. With a professional R&D team, top-notch field application support and rich market experience, World Health enjoys a leading position in the industry in China.






Click "Read original text" to contact us
↓↓↓

Featured Posts


Latest articlesabout

 
EEWorld WeChat Subscription

 
EEWorld WeChat Service Number

 
AutoDevelopers

About Us About Us Service Contact us Device Index Site Map Latest Updates Mobile Version

Site Related: TI Training

Room 1530, Zhongguancun MOOC Times Building,Block B, 18 Zhongguancun Street, Haidian District,Beijing, China Tel:(010)82350740 Postcode:100190

EEWORLD all rights reserved 京B2-20211791 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号 Copyright © 2005-2021 EEWORLD.com.cn, Inc. All rights reserved