Repair and maintenance of flow meter

Publisher:数字冲浪Latest update time:2022-04-20 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

  Maintenance:
  1) Check the sensor power cable and transmission cable (or conductor) frequently for damage or aging, and maintain the rubber sheath on the cable.
  2) Wipe the electrode surface with a soft cloth regularly (usually every six months) to remove dust or deposits.
  3) Remove debris from the upstream straight channel section of the sensor.


  Faults and their elimination:
  1) Liquid does not flow but has output: 1. The signal transmission cable connection between the converter and the sensor is broken; 2. The signal cable to the electrode is broken; 3. The electrode surface is contaminated or the insulation layer is accumulated; 4. The grounding is poor or the circuit is broken.
Solution: 1. Connect the cable properly; 2. Open the sensor and reconnect it; 3. Wipe the electrode surface; 4. Connect the bottom line.
  2) Liquid does not flow but has no output: 1. The two cores of the signal transmission cable between the converter and the sensor are connected in reverse; 2. The power supply is not connected or the contact is poor; 3. The sensor surface pipe, shell, and end face have leakage.
Solution: 1. Reverse the wire head; 2. Connect the power supply and keep the contact good; 3. Repair the sensor.
  3) Unstable output: 1. Unstable flow field; 2. The liquid passing through the sensor contains gas or large solid blocks; 3. Electrical connection is not connected; 4. Poor grounding; 5. Electrode leakage.
  Solution: 1. Modify the path or add a false sensor; 2. Normal phenomenon; 3. Check the wiring and connect the wires; 4. Connect the ground wire; 5. Repair the sensor.
  4) Large error: 1. The zero point is too high; 2. Not completely filled with liquid; 3. The power supply distortion is too large; 4. Poor grounding.
  Solution: 1. Re-adjust the zero point; 2. Improve the path conditions so that the sensor is always filled with liquid; 3. Improve the power supply conditions to meet normal working conditions; 4. Connect the ground wire

Reference address:Repair and maintenance of flow meter

Previous article:How to use electronic insulation megohmmeter
Next article:Structure and principle of 5050 megohmmeter

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号