Tektronix TDS3054C Oscilloscope White Screen Repair Case

Publisher:ShimmeringMoonLatest update time:2022-01-18 Source: eefocusKeywords:Tektronix Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

1. Instrument model: Tektronix TDS3054C digital phosphor oscilloscope;


2. Fault description: The first screen is white after power on;

3. Testing process:

1. Disassemble the oscilloscope, clean the inside of the instrument, re-enter and tighten each cable interface; the fault does not disappear;

2. After powering on, it was found that the parameters of some chips of the instrument were abnormal, and it was preliminarily determined that there were faults in the chips and peripheral devices;

3. Further testing confirmed the damaged chip.


4.Maintenance:

Replace the faulty chip and the machine will start up normally; recalibrate the whole machine and adjust the indicators.


5. Repair result: Repaired.


Keywords:Tektronix Reference address:Tektronix TDS3054C Oscilloscope White Screen Repair Case

Previous article:Application of Tektronix Oscilloscope MSO5 Series and MSO6 Series in RF Pulse Test
Next article:Tektronix DPO4054 oscilloscope self-test failed acquisition failed calibration error fault repair

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号