Keysight/Agilent N5230C network analyzer cannot enter the system failure case

Publisher:创意梦者Latest update time:2021-12-02 Source: eefocusKeywords:Keysight Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

1. Instrument model:

Agilent/Keysight N5230C Microwave Network Analyzer.


2. Fault phenomenon:

After booting up, the computer stays on the test interface and cannot enter the system.

3. Testing process:

The instrument performs internal self-check and finds that the instrument reports an error when entering the measurement interface and cannot enter the test interface normally.


4. Maintenance process:

When a system error occurs and the system cannot be entered normally, first determine whether there is a problem with the system. After testing, it is found that the instrument system and firmware are damaged.

Repair measures: Reinstall the system and restore the firmware.


5. Post-repair inspection:

Power on the machine for self-test, eliminate all error messages, and repair faults.


Keywords:Keysight Reference address:Keysight/Agilent N5230C network analyzer cannot enter the system failure case

Previous article:Agilent HP HP8753 Series Network Analyzer Calibration Parameters Settings
Next article:TYPE576 Crystal Characteristic Graphic Analyzer Maintenance Case

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号