Test steps of synthetic blood penetration tester for medical meltblown filter material

Publisher:RadiantSerenityLatest update time:2022-10-09 Source: elecfans Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

Medical meltblown filter material synthetic blood penetration tester is used to measure the penetration performance of liquid source such as synthetic blood, so as to evaluate its safety protection performance.


Main application: It is suitable for the determination of the penetration performance of synthetic blood splash of medical melt-blown filter materials. It can be used by medical inspection departments, safety inspection departments and scientific research institutions.


The medical mask sample is supported on the test device, and 2mL of synthetic blood is sprayed horizontally from a cannula with an inner diameter of 0.84mm toward the meltblown filter material to be tested at a distance of 300mm. The scene of blood splashing from a perforated blood vessel is simulated, and the penetration of synthetic blood on the other side of the meltblown filter material is observed. The medical mask samples are tested at the liquid spray pressures corresponding to the liquid spray speeds of 450cm/s, 550cm/s, and 635cm/s. The test results of each spray speed are recorded, and the maximum blood pressure corresponding to the acceptable quality level limit of 4 for the medical mask is given.


experiment procedure:

1. Pull the handle of the sample clamp on the upper left of the instrument and pull the sample clamp to the left. Fix the sample (the standard requires that the sample be tested within 60 seconds after being taken out of the pretreatment chamber) on the sample clamp. If the melt-blown filter material has wrinkles, unfold the wrinkles and fix them on the sample clamp to ensure that the target area is a single layer of material. The center of the sample is used as the test target area. Close the sample clamp.

2. Press the "Start" button. At this time, a certain amount of synthetic blood (surface tension (0.042±0.002) N/m) is sprayed horizontally toward the melt-blown filter material behind the positioning target. At the same time, the timer starts timing. After the set spraying time, the instrument automatically stops spraying. After 10 seconds, the buzzer sounds. Press the "Stop" button and check whether there is synthetic blood on the observation surface of the sample.

3. One test is over. If you need to continue the test, repeat steps AC until all tests are completed. After all the tests are completed, clean the synthetic blood residue in the test liquid box, replace it with distilled water, and clean the entire pipeline. When the nozzle sprays out clean water, continue to drain all the clean water in the test liquid tank. Close the outlet valve on the air compressor and turn off the power switch. Wash the residual liquid tray, clean the work surface, and clean the distilled water in the test liquid box.

4. Positioning target test. First, close the outlet valve on the air compressor, connect the air compressor to the power supply, and wait until the air compressor reaches the predetermined pressure. Cut off the power supply of the air compressor, turn off the gas source switch, open the outlet valve, pull up the valve head of the filter pressure reducing valve on the left side of the instrument, and follow the valve head indication. The "+" direction is for pressurization, and the "-" direction is for pressure reduction adjustment, so that the pressure gauge of the filter pressure reducing valve displays 0.2MPa. Install the positioning target so that the distance between the back of the positioning target and the melt-blown filter material sample is 10mm, and tighten the positioning screw on the guide plate at the bottom of the positioning target to position it.

5. Turn the air source switch on the left side of the instrument to off (vertical is off), turn on the power switch, press the "test solution" button, pay attention to the return pipe above the test solution box, and press the "test solution" button immediately when the test solution flows out of the return pipe. The test solution is applied. After using the positioning ruler to determine that the distance from the nozzle to the mask is (300±10) mm, fix the positioning target seat with screws. Unscrew the air source switch on the left side of the instrument (the knob is horizontal), adjust the precision pressure reducing valve on the top of the instrument, follow the valve head indication, the "+" direction is for pressure increase, and the "-" direction is for pressure reduction adjustment, refer to Table 2 to adjust the pressure gauge pressure, set the "spray time" timer on the panel to 1min or longer, press the "start" button, loosen the nozzle positioning screw, adjust the nozzle up, down, left, and right, so that the sprayed test solution flow is aimed at the center of the small hole on the positioning target plate in a straight shot, tighten the positioning screw, and press the "stop" button.


experiment procedure:

Hold the handle of the sample clamp on the upper left of the instrument and pull the sample clamp to the left. Fix the sample (the standard requires that the sample be tested within 60 seconds after being taken out of the pretreatment chamber) on the sample clamp. If the melt-blown filter material has wrinkles, unfold the wrinkles and fix them on the sample clamp to ensure that the target area is a single layer of material. The center of the sample is used as the test target area. Close the sample clamp. Press the "Start" button. At this time, a certain amount of synthetic blood (surface tension (0.042±0.002) N/m) is sprayed horizontally toward the melt-blown filter material to be tested behind the positioning target. The spraying starts, and the time relay starts timing. After the set spraying time, the instrument automatically stops spraying.


After 10 seconds, the buzzer sounds. Press the "Stop" button to open the sample clamp. Under appropriate lighting conditions, check whether synthetic blood appears on the observation surface of the sample. One test is over. If you need to continue the test, repeat steps AC until all tests are completed. After all the tests are completed, clean the synthetic blood residue in the test liquid box, replace it with distilled water, and clean the entire pipeline. When the nozzle sprays clean water, continue to drain all the clean water in the test liquid tank. Close the outlet valve on the air compressor and turn off the power switch. Wash the residual liquid tray, clean the work surface, and clean the distilled water in the test liquid box.


Reference address:Test steps of synthetic blood penetration tester for medical meltblown filter material

Previous article:Battery slurry solid content moisture test steps and precautions
Next article:What is the applicable scope of UL94VTM horizontal vertical combustion tester?

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号