Some people may not understand why the probe should be compensated.
Let me explain in detail, hoping that everyone can understand it thoroughly.
All of this starts with understanding the compensation principle of the probe:
Oscilloscope input resistance
The oscilloscope probe cannot send the circuit signal into the oscilloscope. At first glance, it seems that it can be used by directly connecting it.
However, when we use a multimeter to measure the resistance at both ends of the oscilloscope probe, it is nearly 9M ohms, as shown in the following figure:
If we look at an oscilloscope, careful friends will find that the 1MΩ input resistance parameter to ground is usually marked next to the BNC input interface of the oscilloscope. Many people may not understand what this means.
In fact, when using an oscilloscope probe to measure a circuit, since we do not want the connection of the oscilloscope probe to change the working state of the measured circuit itself, the oscilloscope probe must be high-impedance, that is, the input impedance is relatively large (megaohm level). The oscilloscope has a certain voltage input range, but different measurement occasions will have different voltages, so the oscilloscope probe will have different attenuation ratios (1X, 10X, 100X...). Then the simplest signal attenuation is achieved through resistor voltage division, as shown in the following figure:
In the figure, R1 is the resistance on the oscilloscope probe, and Rin is the input resistance of the oscilloscope. Generally, Rin = 1MΩ, R1 = 99MΩ at 100X, R1 = 9MΩ at 10X, and theoretically it should be 0Ω at 1X, but in fact R1 is about several hundred ohms, generally within 300 ohms.
Oscilloscope input capacitance
So can the oscilloscope be used according to the resistor voltage divider circuit introduced above? No.
As we all know, in reality, no circuit is an ideal circuit and has more or less parasitic parameters. The interface between the oscilloscope and the oscilloscope probe is no exception. Since the oscilloscope interface needs to connect the signal and GND to the oscilloscope probe at the same time (as shown in the figure below, the metal on the outer circle is generally GND, which can play a role of shielding with the outside, and the metal inside is the input signal), a capacitor is formed between the input signal and GND. No matter how the design of the oscilloscope interface is improved, the parasitic parameters of the oscilloscope input capacitor cannot be eliminated.
The typical values of input capacitance of general oscilloscopes are 15pF, 14pF, and 12pF. The one shown in the figure is 14pF.
There are both R and C, isn’t this an RC low-pass filter?
Let's calculate the cutoff frequency of this RC circuit. Considering the 10X gear, R1 = 9MΩ, Rin = 1MΩ, Cin = 14pF, the cutoff frequency is
In this way, all signals higher than 12.64kHz are attenuated to the point where they are unreadable. So how do we solve this problem? Reduce Cin? Impossible, because physical limitations dictate that Cin must exist, and 14pF itself is already a very small capacitance value; reduce R1 and Rin? Too small a resistance will inevitably affect the measured circuit. It seems that there is no other way. But there are always smart people who can find a solution: compensation capacitor
As long as it satisfies
This allows signals to be transmitted at the correct 10:1 attenuation ratio at relatively high frequencies without causing signal distortion.
But a new problem arises. Different oscilloscopes have different Cin. Even if the model is the same, different oscilloscopes may not meet the above ratio when using the same probe due to inconsistent manufacturing parameters. Doesn't that mean that a specific probe must be manufactured for each oscilloscope? How to solve the universality problem? It's very simple, just add another variable capacitor. As shown in the figure Cp.
In this way, the problem can be solved by simply adjusting the size of the variable capacitor on the probe and observing the waveform distortion on the oscilloscope.
The oscilloscope generally outputs a 1KHz, 5V (or less) square wave signal, which is used for probe compensation calibration. This signal is often indicated by a square wave symbol plus a ground symbol. We can use this signal as a signal source. Start the oscilloscope, as shown in the figure below, connect the probe's BNC to channel 1, and the other end to the square wave signal output port.
When the following two situations occur, it means that the probe compensation is incorrect and you need to use the "adjustment stick" to adjust the compensation capacitor on the probe.
Use the adjustment rod to turn the screw in the probe screw hole to adjust the compensation capacitor to obtain the correct waveform.
Some probes also set it at one end of the detection head, as shown below:
Therefore, when the oscilloscope is replaced with a new probe or the probe has not been used for a long time, we should perform compensation calibration on the probe.
Previous article:How to measure power supply ripple with an oscilloscope
Next article:Capture and analyze LIN protocol waveform using an oscilloscope
- Popular Resources
- Popular amplifiers
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
- In what situations are non-contact temperature sensors widely used?
- How non-contact temperature sensors measure internal temperature
- LED chemical incompatibility test to see which chemicals LEDs can be used with
- Application of ARM9 hardware coprocessor on WinCE embedded motherboard
- What are the key points for selecting rotor flowmeter?
- LM317 high power charger circuit
- A brief analysis of Embest's application and development of embedded medical devices
- Single-phase RC protection circuit
- stm32 PVD programmable voltage monitor
- Introduction and measurement of edge trigger and level trigger of 51 single chip microcomputer
- Improved design of Linux system software shell protection technology
- What to do if the ABB robot protection device stops
- CGD and Qorvo to jointly revolutionize motor control solutions
- CGD and Qorvo to jointly revolutionize motor control solutions
- Keysight Technologies FieldFox handheld analyzer with VDI spread spectrum module to achieve millimeter wave analysis function
- Infineon's PASCO2V15 XENSIV PAS CO2 5V Sensor Now Available at Mouser for Accurate CO2 Level Measurement
- Advanced gameplay, Harting takes your PCB board connection to a new level!
- Advanced gameplay, Harting takes your PCB board connection to a new level!
- A new chapter in Great Wall Motors R&D: solid-state battery technology leads the future
- Naxin Micro provides full-scenario GaN driver IC solutions
- Interpreting Huawei’s new solid-state battery patent, will it challenge CATL in 2030?
- Are pure electric/plug-in hybrid vehicles going crazy? A Chinese company has launched the world's first -40℃ dischargeable hybrid battery that is not afraid of cold
- For electronic hardware, there are many circuit structures, which can be said to be the units that make up a component. All circuits...
- TDA2030 waveform output is wrong
- Schematic diagram - how does DZ1 protect the MOS tube?
- Some records of 4680 high nickel battery research
- 【GD32450I-EVAL】+11 function integration
- Full-bridge converter circuit
- What does the accuracy and resolution of a servo encoder affect?
- MSP430G2553 clock system configuration considerations
- NB-IoT is very popular, what can it do?
- [CNC adjustable power supply] Based on 7508DK- Coding work progress update