Precautions before using the network analyzer

Publisher:机器人总动员Latest update time:2021-11-29 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

Network analyzers are the main instruments for testing and debugging RF components. With the rapid development of radio communication equipment, the demand for RF devices has also increased dramatically, and network analyzers are one of the indispensable instruments.


Today, Antai Testing will simply share with you some precautions before using the network analyzer to help you use the network analyzer correctly. After all, the price of a mid-to-high-end network analyzer from an imported brand is similar to that of a car. If the instrument is damaged due to improper operation, it would be a loss.

The basic function of a vector network analyzer is S parameter testing. The so-called S parameters are scattering parameters, which describe the incident wave quantity, reflected wave quantity, and wave quantity transmission characteristics of electromagnetic waves in the device under test. Simply put: S11 represents the reflection of port 1, S22 represents the reflection of port 2, S21 is the transmission from port 1 to port 2, and S12 is the transmission from port 2 to port 1.

Setup before testing:

§ Network analyzer port connected to a dedicated test cable;

§ The frequency range is set according to the frequency range of the device under test (DUT);

§ When measuring a DUT such as an amplifier with maximum gain Gain, set the output power PWR>>Power: -Gain. Also note that the DUT output power must not exceed the range (such as 0dBm);

§ IF bandwidth setting is based on test standard or BW>>Bandwidth: 1kHz;

§ The number of measurement points is based on standard or Sweep>>number of points: 401;

§ Connect the auto-calibration device to perform calibration CAL>>Start Autocal;

§ If there is only a manual calibration part, the vector network must load the data file that matches the calibration part, and ideal data cannot be used to perform UOSM or TOSM calibration;

§ Note that the type of the test port of the dedicated test cable must be consistent with the calibration component and cannot be transferred.

After the above settings and calibration, select the required test items for testing.

Reference address:Precautions before using the network analyzer

Previous article:Interpretation of key technical indicators of vector network analyzer
Next article:Application of Rohde Network Analyzer ZVL in RF Product Measurement

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号