Analysis and application of probe bias capability and dynamic range

Publisher:asa1670Latest update time:2020-02-12 Source: elecfansKeywords:Probe Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

When using active probes for testing, the most common mistake is using the probe beyond its dynamic range. We often receive calls from customers saying that when they use active and passive probes to measure the same signal, the voltage of the active probe is lower than that of the passive probe. The reason for this is that in most cases it is caused by a misunderstanding of the dynamic range and the withstand voltage range.

Taking the Keysight N2795A active probe as an example, its withstand voltage range is plus or minus 20V, input dynamic range is plus or minus 8V, and offset range is plus or minus 8V:

Analysis and application of probe bias capability and dynamic range

>>> Input dynamic range is defined as

The input dynamic range refers to the voltage range above and below the center line of the oscilloscope screen that the probe can test. For example, a probe with a ±2.5V dynamic input range can only measure voltages within 2.5V above and below the center line of the oscilloscope screen. If the input signal fluctuates beyond this range, it will be reflected in the measured waveform as clipping, and the measured amplitude will be smaller.

By definition, this means that when using the N2795A probe, you can measure waveforms within 8V above and below the center line of the oscilloscope screen:

Analysis and application of probe bias capability and dynamic range

When we directly measure the 0-16V sine wave, since the waveform exceeds the range of 8V above the center line of the screen, it will cause waveform distortion and make the test result low.

Analysis and application of probe bias capability and dynamic range

This is where the probe/oscilloscope bias capability comes into play:

>>> The definition of bias capability is:

Bias capability refers to the ability to adjust the 0V voltage reference line to the voltage difference with the center line of the oscilloscope screen. By setting a suitable bias according to the DC component of the signal, the dynamic signal with a DC component can be adjusted to the vicinity of the center line of the oscilloscope screen to meet the requirements of the dynamic input range of the probe. 

For example, if the distorted test waveform above is set to an 8V bias so that the waveform continues to be displayed in the center of the screen, the 0-16V sine wave voltage can be tested correctly:

Analysis and application of probe bias capability and dynamic range

Therefore, the maximum input voltage of the probe is not necessarily the measurable voltage range. The measurable voltage range should be its dynamic range plus the offset range. Of course, the offset setting should be used reasonably at this time so that the test waveform is always displayed within its dynamic range.

Keywords:Probe Reference address:Analysis and application of probe bias capability and dynamic range

Previous article:Introduction and analysis of VB programming for oscilloscope remote control
Next article:Analysis and introduction of probe parameters

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号