An article explains the difference between zero calibration and zero elimination in power analyzers

Publisher:创意狂想Latest update time:2020-01-11 Source: elecfans Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

1. Zero calibration

Let's first explain the impact of zero-point error on normal measurement. The basic measurement circuit is shown in the figure below:

An article explains the difference between zero calibration and zero elimination in power analyzers

Figure 1 Basic measurement circuit diagram for zero calibration

Suppose the actual external input signal is , for the op amp conditioning circuit, the zero offset voltage will be amplified along with the signal, and the zero point error is , as shown in Figure 1, then the actual input to the amplifier circuit is , and the signal actually input to the ADC after amplification is .

Zero point error is superimposed on the actual input signal to cause measurement error. Generally, the instrument will be zero-calibrated before leaving the factory, but the zero point will drift due to the influence of the use environment and device aging. Therefore, zero point calibration can also be enabled when the customer uses it. The principle is shown in the figure below:

An article explains the difference between zero calibration and zero elimination in power analyzers

Figure 2 Enable calibration circuit diagram


During zero calibration, SW1 is disconnected and SW2 is closed. The signal collected by the ADC is : During normal sampling, SW2 is disconnected and SW1 is closed. The signal collected by the ADC is :

At this time, the signal collected is the actual desired signal, the zero voltage is eliminated, the accuracy is higher, but it has no effect on the actual input signal. The PA series supports zero calibration function, as shown in Figure 3, you can click the "zero calibration" button to manually calibrate the zero or turn on "auto zero calibration". When the range, filter and other settings are changed, automatic re-zeroing is required.

An article explains the difference between zero calibration and zero elimination in power analyzers

Figure 3 Calibration instrument interface settings


2. Zero elimination

Zero elimination is the NULL function of PA. The NULL function does not change the actual input signal, but only performs arithmetic addition and subtraction on the collected signal to achieve translation, which is similar to providing the user with an arithmetic zero point to replace the actual signal zero point. As shown in Figure 4, measure the sine signal superimposed on the Vdc DC, connect the normal signal to the instrument and click the "NULL" button, then the DC signal Vdc will be used as the new zero point value, and the sine signal centered at 0 will be presented to the user after sampling, and Vdc will be eliminated. At this time, if the DC signal connected to Vdc will show that the measured signal is 0.

An article explains the difference between zero calibration and zero elimination in power analyzers

Figure 4 Schematic diagram of zero elimination


The NULL function is only suitable for specific users to process some specific signals. The signal presented to the user is different from the actual signal. Turning on the NULL function in normal use will cause measurement errors. The PA series power analyzer supports the NULL function. The setting diagram is as follows. It also supports users to enter a custom NULL value, and the set offset value is subtracted from the DC component of the measurement loop.

An article explains the difference between zero calibration and zero elimination in power analyzers

Figure 5 Instrument NULL setting menu


In summary, the zero calibration function is used by the instrument to eliminate the zero point deviation of the internal measurement circuit and improve the measurement accuracy. The collected signal is still the original signal. The NULL function is that the instrument provides an arithmetic zero point to shift the signal as a whole. The collected signal has a fixed DC deviation from the original signal, which is suitable for specific applications.

Reference address:An article explains the difference between zero calibration and zero elimination in power analyzers

Previous article:Principles and Applications of Digital Oscilloscopes and Logic Analyzers
Next article:Current status of motor start-up test! Difficulties of testing motor start-stop with power analyzer

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号