1. Why use a non-contact infrared thermometer? Non-contact infrared thermometers use infrared technology to quickly and easily measure the surface temperature of an object.
No mechanical contact with the object is required to quickly measure the temperature reading.
Just aim, press the trigger, and read the temperature data on the LCD display.
Infrared thermometers are light, small, easy to use, and can reliably measure hot, dangerous or hard-to-reach objects without contaminating or damaging the object.
Infrared thermometers can measure several readings per second, while contact thermometers take several minutes to measure per second.
2. How does infrared work? Infrared thermometers receive invisible infrared energy emitted by various objects themselves. Infrared radiation is part of the electromagnetic spectrum, which includes radio waves, microwaves, visible light, ultraviolet rays, R rays and X rays.
Infrared is located between visible light and radio waves. Infrared wavelengths are often expressed in microns, with a wavelength range of 0.7 microns to 1000 microns. In fact, the 0.7 micron to 14 micron band is used for infrared temperature measurement.
3. How to ensure temperature measurement accuracy? The undisputed understanding of infrared technology and its principles is its accurate temperature measurement.
When the temperature is measured by a non-contact instrument, the infrared energy emitted by the object being measured is converted into an electrical signal on the detector through the optical system of the thermometer, and the temperature reading of the signal is displayed. There are several important factors that determine the accurate temperature measurement. The most important factors are emissivity, field of view, distance to the light spot and the position of the light spot.
Emissivity, all objects reflect, transmit and emit energy, and only the emitted energy can indicate the temperature of the object.
When the infrared thermometer measures the surface temperature, the instrument can receive all three types of energy.
Therefore, all thermometers must be adjusted to read only the emitted energy.
Measurement errors are usually caused by infrared energy reflected from other light sources.
Some measuring instruments can change the emissivity, and the emissivity values of various materials can be found in published emissivity tables.
Other instruments are fixed and preset to an emissivity of 0.95.
This emissivity value is for the surface temperature of most organic materials, paints or oxidized surfaces. It is necessary to apply a tape or flat black paint to the measured surface to compensate.
When the tape or paint reaches the same temperature as the base material, the temperature of the tape or paint surface is measured, which is its true temperature.
Distance to spot ratio, the optical system of the infrared thermometer collects energy from the circular measurement spot and focuses it on the detector. Optical resolution is defined as the ratio of the distance from the instrument to the object to the size of the measured spot (D:S).
The larger the ratio, the better the resolution of the instrument and the smaller the measured spot size.
Laser aiming, only used to help aim at the measurement point.
The latest improvement in infrared optics is the addition of a close focus feature, which can provide accurate measurement of small target areas and prevent the influence of background temperature.
Field of view, make sure the target is larger than the spot size when the instrument is measuring. The smaller the target, the closer it should be.
When accuracy is particularly important, make sure the target is at least twice the spot size.
4. How to measure temperature? To measure temperature, point the instrument at the object to be measured, press the trigger to read the temperature data on the instrument's LCD, and make sure to arrange the distance to spot size ratio and field of view.
There are several important things to remember when using an infrared thermometer: 1. Only measure surface temperature. Infrared thermometers cannot measure internal temperatures.
2. Temperature measurement cannot be performed through glass. Glass has very specific reflection and transmission characteristics that do not allow accurate infrared temperature readings.
However, temperature measurement can be performed through the infrared window.
Infrared thermometers are best not used to measure the temperature of shiny or polished metal surfaces (stainless steel, aluminum, etc.).
3. Locating hot spots. To find hot spots, aim the instrument at the target and then make an up-and-down scanning motion on the target until the hot spot is located.
4. Pay attention to environmental conditions: steam, dust, smoke, etc.
It blocks the optical system of the instrument and affects accurate temperature measurement.
5. Ambient temperature. If the thermometer is suddenly exposed to an ambient temperature difference of 20 degrees or more, allow the instrument to adjust to the new ambient temperature within 20 minutes.
V. What are the most common applications? There are many applications for non-contact thermometers. The most common ones are: 1. Automotive industry: Diagnosing cylinders and heating/cooling systems.
2. HVAC: Monitoring air stratification, supply/return records, furnace performance.
3. Electrical: Checking faulty transformers, electrical panels and connectors.
4. Food: Scanning management, service and storage temperatures.
5. Others: Many projects, bases and renovation applications.
Reference address:Infrared temperature measurement technology questions and answers
No mechanical contact with the object is required to quickly measure the temperature reading.
Just aim, press the trigger, and read the temperature data on the LCD display.
Infrared thermometers are light, small, easy to use, and can reliably measure hot, dangerous or hard-to-reach objects without contaminating or damaging the object.
Infrared thermometers can measure several readings per second, while contact thermometers take several minutes to measure per second.
2. How does infrared work? Infrared thermometers receive invisible infrared energy emitted by various objects themselves. Infrared radiation is part of the electromagnetic spectrum, which includes radio waves, microwaves, visible light, ultraviolet rays, R rays and X rays.
Infrared is located between visible light and radio waves. Infrared wavelengths are often expressed in microns, with a wavelength range of 0.7 microns to 1000 microns. In fact, the 0.7 micron to 14 micron band is used for infrared temperature measurement.
3. How to ensure temperature measurement accuracy? The undisputed understanding of infrared technology and its principles is its accurate temperature measurement.
When the temperature is measured by a non-contact instrument, the infrared energy emitted by the object being measured is converted into an electrical signal on the detector through the optical system of the thermometer, and the temperature reading of the signal is displayed. There are several important factors that determine the accurate temperature measurement. The most important factors are emissivity, field of view, distance to the light spot and the position of the light spot.
Emissivity, all objects reflect, transmit and emit energy, and only the emitted energy can indicate the temperature of the object.
When the infrared thermometer measures the surface temperature, the instrument can receive all three types of energy.
Therefore, all thermometers must be adjusted to read only the emitted energy.
Measurement errors are usually caused by infrared energy reflected from other light sources.
Some measuring instruments can change the emissivity, and the emissivity values of various materials can be found in published emissivity tables.
Other instruments are fixed and preset to an emissivity of 0.95.
This emissivity value is for the surface temperature of most organic materials, paints or oxidized surfaces. It is necessary to apply a tape or flat black paint to the measured surface to compensate.
When the tape or paint reaches the same temperature as the base material, the temperature of the tape or paint surface is measured, which is its true temperature.
Distance to spot ratio, the optical system of the infrared thermometer collects energy from the circular measurement spot and focuses it on the detector. Optical resolution is defined as the ratio of the distance from the instrument to the object to the size of the measured spot (D:S).
The larger the ratio, the better the resolution of the instrument and the smaller the measured spot size.
Laser aiming, only used to help aim at the measurement point.
The latest improvement in infrared optics is the addition of a close focus feature, which can provide accurate measurement of small target areas and prevent the influence of background temperature.
Field of view, make sure the target is larger than the spot size when the instrument is measuring. The smaller the target, the closer it should be.
When accuracy is particularly important, make sure the target is at least twice the spot size.
4. How to measure temperature? To measure temperature, point the instrument at the object to be measured, press the trigger to read the temperature data on the instrument's LCD, and make sure to arrange the distance to spot size ratio and field of view.
There are several important things to remember when using an infrared thermometer: 1. Only measure surface temperature. Infrared thermometers cannot measure internal temperatures.
2. Temperature measurement cannot be performed through glass. Glass has very specific reflection and transmission characteristics that do not allow accurate infrared temperature readings.
However, temperature measurement can be performed through the infrared window.
Infrared thermometers are best not used to measure the temperature of shiny or polished metal surfaces (stainless steel, aluminum, etc.).
3. Locating hot spots. To find hot spots, aim the instrument at the target and then make an up-and-down scanning motion on the target until the hot spot is located.
4. Pay attention to environmental conditions: steam, dust, smoke, etc.
It blocks the optical system of the instrument and affects accurate temperature measurement.
5. Ambient temperature. If the thermometer is suddenly exposed to an ambient temperature difference of 20 degrees or more, allow the instrument to adjust to the new ambient temperature within 20 minutes.
V. What are the most common applications? There are many applications for non-contact thermometers. The most common ones are: 1. Automotive industry: Diagnosing cylinders and heating/cooling systems.
2. HVAC: Monitoring air stratification, supply/return records, furnace performance.
3. Electrical: Checking faulty transformers, electrical panels and connectors.
4. Food: Scanning management, service and storage temperatures.
5. Others: Many projects, bases and renovation applications.
Previous article:What is TDR? Introduction to TDR Principle
Next article:Key considerations for test and measurement system requirements and instrument selection
Recommended Content
Latest Test Measurement Articles
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
MoreSelected Circuit Diagrams
MorePopular Articles
- LED chemical incompatibility test to see which chemicals LEDs can be used with
- Application of ARM9 hardware coprocessor on WinCE embedded motherboard
- What are the key points for selecting rotor flowmeter?
- LM317 high power charger circuit
- A brief analysis of Embest's application and development of embedded medical devices
- Single-phase RC protection circuit
- stm32 PVD programmable voltage monitor
- Introduction and measurement of edge trigger and level trigger of 51 single chip microcomputer
- Improved design of Linux system software shell protection technology
- What to do if the ABB robot protection device stops
MoreDaily News
- Allegro MicroSystems Introduces Advanced Magnetic and Inductive Position Sensing Solutions at Electronica 2024
- Car key in the left hand, liveness detection radar in the right hand, UWB is imperative for cars!
- After a decade of rapid development, domestic CIS has entered the market
- Aegis Dagger Battery + Thor EM-i Super Hybrid, Geely New Energy has thrown out two "king bombs"
- A brief discussion on functional safety - fault, error, and failure
- In the smart car 2.0 cycle, these core industry chains are facing major opportunities!
- The United States and Japan are developing new batteries. CATL faces challenges? How should China's new energy battery industry respond?
- Murata launches high-precision 6-axis inertial sensor for automobiles
- Ford patents pre-charge alarm to help save costs and respond to emergencies
- New real-time microcontroller system from Texas Instruments enables smarter processing in automotive and industrial applications
Guess you like
- Microcomputer Principles and Interface Design
- What is the relationship between the machine cycle and clock cycle of RH850?
- The highest increase is 30%, 10 IC manufacturers including ST, MediaTek, and ON Semiconductor have increased prices!
- Inductor Current Measurement in Switching Power Supplies
- Xunwei i.MX6ULL terminator process basic exec function family
- LLAKG: Arduino automatic watering system (Episode 1: System concept and function introduction)
- Accelerating the design of electric masks
- Error of PMSM sliding mode observer?
- Please recommend a good digital circuit simulation software
- Ice and Fire Mini Fireplace Made by Seeeduino XIAO