MAINS FILTERS
Medical IEC Power Inlets Flange and Snap Fit
PANEL MOUNTING
1, 3, 6 or 10 Amp Current
Rating
Medical
3 Alternative Circuits
Optional Additional
Components
2.8 or 6.3mm tabs
Flange
PS00/B
SNAP FIT
Snap Fit
PS01/B
1, 3, 6 or 10 Amp Current
Rating
Medical
3 Alternative Circuits
Optional Additional
Components
2.8 or 6.3mm tabs
1, 1.5, 2 or 3mm panels
PS00 or PS01/Bxx x x /xx xx
IEC CONNECTORS
Series
PS00/B
PS01/B
Rating
01 = 1A
03 = 3A
06 = 6A
10 = 10A
L/C Circuit
1 = Version 1
2 = Version 2
3 = Version 3
Additional
Components
0 = None
1 = Bleed (R) Resistor
2 = Surge (VDR) Protection
3 = “R” plus “VDR”
Tag Type and
Configuration
28 = 2.8mm
Panel
Thickness
00 = Flange
63 = 6.3mm tabs 10 = 1.0mm
15 = 1.5mm
20 = 2.0mm
30 = 3.0mm
Specification
Max. Working Voltage:
Earth Leakage Current:
Temperature Range:
Max. Ambient Temp.:
(@ Full Load)
Test Voltage:
Approvals:
PS00/Bxxxx/xx00
250V a.c. 50-400Hz
<100µA (typically 5µA 250V, 50Hz)
–25ºC to +85ºC
40ºC (derate linearly to 0A @ 85ºC)
2700V d.c. 2 secs. Lines to Earth
1100V d.c. 2 secs. Live to Neutral
PS01/Bxxxx/xxxx
250V a.c. 50-400Hz
<100µA (typically 5µA 250V, 50Hz)
–25ºC to +85ºC
40ºC (derate linearly to 0A @ 85ºC)
2700V d.c. 2 secs. Lines to Earth
1100V d.c. 2 secs. Live to Neutral
Part No. Example
PS00/B0120/6300 =
PS00 series, flange fitting, filtered
IEC power inlet, for medical
applications, rated at 1 ampere.
L/C circuit version 2, i.e. L1 = 2
x 10mH, Cx = 15nF, without
bleed resistor or surge protection
device fitted, with 6.3mm tabs.
Mating Connectors:
PX0587, PX0587/SE, PX0588
Compliant
PX0587, PX0587/SE, PX0588
Compliant
RoHS
134
www.elektron-bulgin.com