The purpose of this article is to provide you with a basic overview of a spectrum analyzer or signal analyzer. You may want to learn more about other topics related to spectrum analysis, you can visit the spectrum analyzer web page. Here we will focus on the basic principles of spectrum analyzer operation. Although today's technology has made it possible to replace many analog circuits with modern digital implementations, it is still very beneficial to start with the classic spectrum analyzer architecture. In the future we will also look at the capabilities and advantages that digital circuitry brings to spectrum analyzers, and discuss the digital architecture used in modern spectrum analyzers.
Figure 2-1 is a simplified block diagram of a superheterodyne spectrum analyzer. "Heterodyne" refers to mixing, which is the conversion of frequencies, while "hyper" refers to superaudio frequencies or the frequency range above audio. From the figure we see that the input signal first passes through an attenuator, then a low-pass filter (you will see why the filter is placed here later), reaches the mixer, and then is combined with the signal from the local oscillator (LO). phase mixing.
Figure 2-1. Block diagram of a typical superheterodyne spectrum analyzer
Since the mixer is a nonlinear device, its output contains not only the two original signals, but also their harmonics, as well as the sum and difference signals of the original signal and its harmonics. If any of the mixed signals fall within the passband of the intermediate frequency (IF) filter, it is further processed (amplified and possibly logarithmically compressed). The basic processing procedures include envelope detection, low-pass filtering and display. The ramp generator produces horizontal movement from left to right on the screen, and it also tunes the local oscillator so that the change in local oscillator frequency is proportional to the ramp voltage.
If you are familiar with a superheterodyne AM radio that receives ordinary amplitude modulation (AM) broadcast signals, you will find that its structure is very similar to the block diagram shown in Figure 2-1. The difference is that the output of the spectrum analyzer is a screen rather than a speaker, and its local oscillator is tuned electronically rather than by a front panel knob.
Since the output of the spectrum analyzer is the XY trace on the screen, let's see what information we can get from it. The display is mapped on a dial consisting of 10 horizontal grids and 10 vertical grids. The horizontal axis represents frequency, with scale values increasing linearly from left to right. Frequency setting is usually divided into two steps: first adjust the frequency to the center line of the dial through the center frequency control, and then adjust the frequency range (span) across 10 grids through the frequency span control. These two controls are independent of each other, so when you change the center frequency, the span does not change. Also, we can set the start frequency and stop frequency instead of setting the center frequency and span. In either case, we can determine the absolute frequency of any displayed signal and the relative frequency difference between any two signals.
The vertical axis scale is divided by amplitude. You can choose a linear scale scaled in voltage or a logarithmic scale scaled in decibels (dB). The logarithmic scale is more often used than the linear scale because it reflects a larger range of values. The logarithmic scale can simultaneously display signals whose amplitudes differ by 70 to 100 dB (voltage ratios from 3200 to 100,000 or power ratios from 10,000,000 to 10,000,000,000), while the linear scale can only be used with amplitudes that differ by no more than 20 to 30 dB (voltage ratios). than 10 to 32) signal. In both cases, we will use calibration technology 1 to give the level of the highest row on the dial, that is, the absolute value of the reference level, and determine other positions on the dial based on the proportion corresponding to each small cell. value. In this way, we can measure both the absolute value of the signal and the relative amplitude difference of any two signals.
The scale values for frequency and amplitude are annotated on the screen. Figure 2-2 is a typical spectrum analyzer display.
Figure 2-2. Typical spectrum analyzer display with parameters set
Now let's return our attention to the spectrum analyzer components shown in Figure 2-1.
RF attenuator
The first part of the analyzer is the RF attenuator. Its role is to ensure that the signal is at the appropriate level when entering the mixer, thereby preventing overload, gain compression and distortion. Because the attenuator is a protection circuit for the spectrum analyzer, it is usually set automatically based on a reference level value, but the attenuation value can also be manually selected in steps of 10 dB, 5 dB, 2 dB, or even 1 dB. Figure 2-3 shows an example of an attenuator circuit with a maximum attenuation value of 70 dB in 2 dB steps.
The DC blocking capacitor is used to prevent the analyzer from being damaged by DC signals or DC bias of signals. However, it will attenuate low-frequency signals and increase the lowest available starting frequency of some spectrum analyzers to 9 kHz and 100 kHz. or 10 MHz.
In some analyzers, an amplitude reference signal can be connected as shown in Figure 2-3, which provides a signal with precise frequency and amplitude for periodic self-calibration of the analyzer.
Figure 2-3. RF attenuator circuit
Low pass filter or preselector
The purpose of the low-pass filter is to prevent high-frequency signals from reaching the mixer. This prevents out-of-band signals from mixing with the local oscillator and producing unwanted frequency response at the intermediate frequency. Microwave spectrum analyzers or signal analyzers replace low-pass filters with preselectors, which are adjustable filters that filter out signals at frequencies other than those of interest. In Chapter 7, we will introduce in detail the purpose and method of filtering input signals.
Analyzer tuning
We need to know how to tune a spectrum analyzer or signal analyzer to the frequency range we want. Tuning depends on the center frequency of the IF filter, the frequency range of the local oscillator, and the frequency range that allows external signals to reach the mixer (allowing them to pass through the low-pass filter). Of all the signal components output from the mixer, there are two signals with the largest amplitude that we are most interested in. They are the signal components produced by the sum of the local oscillator and the input signal and the difference between the local oscillator and the input signal. If we can make the signal we want to observe an IF higher or lower than the local oscillator frequency, then one of the desired mixing components will fall within the passband of the IF filter and will then be detected and produce an amplitude on the screen. response.
In order to tune the analyzer to the desired spectral range, we need to select the appropriate local oscillator frequency and intermediate frequency. Assuming that the required tuning range is 0 to 3.6 GHz, the next step is to select the IF frequency. If we choose an IF of 1 GHz, which is within the desired tuning range, we get a 1 GHz input signal, and since the output of the mixer contains the original input signal, the 1 GHz input from the mixer The signal will have a constant output at the intermediate frequency. So no matter how the local oscillator is tuned, a 1 GHz signal will pass through the system and give a constant amplitude response on the screen. The result is a blank area within the frequency tuning range that cannot be measured because the signal amplitude response in this area is independent of the local oscillator frequency. Therefore, the IF frequency of 1 GHz cannot be selected.
That is, we should select the mid-frequency at a higher frequency than the tuning band. In Keysight's X-Series signal analyzers, which are tunable to 3.6 GHz, the first local oscillator frequency range is 3.8 to 8.7 GHz, and the selected IF frequency is approximately 5.1 GHz.
Now we want to tune from 0 Hz (since the instrument with this structure cannot observe the 0 Hz signal, it is actually from a low frequency) to 3.6 GHz.
By choosing the local oscillator frequency to start at the intermediate frequency (LO - IF = 0 Hz) and tune upward to 3.6 GHz above the intermediate frequency, the LO - IF mixing component can cover the required tuning range. Using this principle, the following tuning equation can be established:
If you want to determine the local oscillator frequency required to tune the analyzer to a low, mid, or high frequency signal (such as 1 kHz, 1.5 GHz, or 3 GHz), you first transform the tuning equation to obtain fLO:
Figure 2-4. To produce a response on the display, the local oscillator must be tuned to fIF + fs
Figure 2-4 illustrates the analyzer tuning process. In the figure, fLO is not high enough to cause the fLO -fsig mixing component to fall within the IF passband, so there is no response on the display. However, if the ramp generator is adjusted so that the local oscillator is tuned to a higher frequency, then at some point on the ramp (sweep) the mixing product will fall within the IF passband and we will see a response on the display.
Because the ramp generator simultaneously controls the horizontal position of the trace on the display and the local oscillator frequency, the horizontal axis of the display can be calibrated based on the frequency of the input signal.
We haven't quite solved the tuning problem yet. What happens if the input signal frequency is 9.0 GHz? When the local oscillator is tuned in the 3.8 to 8.7 GHz range, when it reaches the intermediate frequency (3.9 GHz) away from the 9.0 GHz input signal, it gets a mixing product with a frequency equal to the intermediate frequency and generates a response on the display. In other words, the tuning equation easily becomes:
Previous article:On-site inspection case of RSE60 online thermal imaging camera detecting micron-level small targets
Next article:Factors affecting the measurement accuracy of infrared thermometer
- Popular Resources
- Popular amplifiers
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
- In what situations are non-contact temperature sensors widely used?
- LED chemical incompatibility test to see which chemicals LEDs can be used with
- Application of ARM9 hardware coprocessor on WinCE embedded motherboard
- What are the key points for selecting rotor flowmeter?
- LM317 high power charger circuit
- A brief analysis of Embest's application and development of embedded medical devices
- Single-phase RC protection circuit
- stm32 PVD programmable voltage monitor
- Introduction and measurement of edge trigger and level trigger of 51 single chip microcomputer
- Improved design of Linux system software shell protection technology
- What to do if the ABB robot protection device stops
- Huawei's Strategic Department Director Gai Gang: The cumulative installed base of open source Euler operating system exceeds 10 million sets
- Download from the Internet--ARM Getting Started Notes
- Learn ARM development(22)
- Learn ARM development(21)
- Learn ARM development(20)
- Learn ARM development(19)
- Learn ARM development(14)
- Learn ARM development(15)
- Analysis of the application of several common contact parts in high-voltage connectors of new energy vehicles
- Wiring harness durability test and contact voltage drop test method
- When connecting the STEVAL-IDB008V2 board to BlueNRG GUI, the device cannot be found.
- Happy holidays to everyone. During the holidays, goldfish eat meat and stay at home. Here are the statistics
- Discussion: 5G is coming, what skills do we need to acquire? (Give away Chip Coins)
- A LiDAR Level Meter Silo Level Measurement Solution
- Multisim simulation IR2103
- 7 common 51 single-chip microcomputer clock circuit diagrams, collect them!
- [Ebit E840-DTU Review] Performance Introduction
- Design of DCDC voltage stabilization power supply and peripheral circuit for voltage stabilization chip selection
- [Repost] Enter the top ten most commonly used electronic components, how many do you know? (I)
- Help, which chip can be found in TINA TI simulation software to make APFC switching power supply?