Calibration method of withstand voltage tester
1. Tester output voltage
Figure 5 Output voltage calibration wiring diagram
Put the withstand voltage tester in the "reset" state, turn the "voltage adjustment" knob counterclockwise to the bottom, and connect the wires according to Figure 5. Press the "start" button, adjust the "voltage adjustment" button, so that the reading of the high-voltage meter is the value of the detection point shown in Table 3, and calculate whether the error between the indicated value and the reading of the high-voltage meter meets the technical indicators. If not, adjust the corresponding potentiometer value.
table 3
Note: Different output voltages should be selected with a suitable high-voltage meter’s measuring range. The error of the high-voltage meter should be within ±1.5%. If it is a pointer-type high-voltage meter, its measuring range should be above 1/3 of the scale.
2. Leakage current and alarm current
Figure 6 Leakage current and alarm current wiring diagram
Put the withstand voltage tester in the "reset" state, turn the "voltage adjustment" knob counterclockwise to the bottom, connect according to Figure 6, and set the "leakage current range" switch to 2mA. Select the corresponding load resistance according to Table 4.
Table 4
Common faults and troubleshooting methods of withstand voltage tester
1. No power indication when starting up, and the display is off
1. Please check whether the power supply is normal. If not, please check whether there is 220V and whether it is within the normal use range of the instrument;
2. Please check whether the fuse is damaged. If the fuse is damaged, please replace it with the same type of fuse;
3. Please check whether the power transformer has output. If there is no output, please replace the power transformer;
4. Please check whether the voltage regulator circuit 7812 is damaged. If it is damaged, please replace 7812.
2. Start-up alarm:
1. Please check whether the sampling resistor on the leakage current switch board is damaged. If damaged, please replace the sampling resistor of the corresponding gear;
2. Please check whether LM324 is damaged. If it is damaged, please replace LM324.
3. Voltage indication when power on
1. Please check whether the thyristor is damaged. If it is damaged, please replace it with the same specification thyristor;
2. Please check whether the optocoupler 3061 is damaged. If it is damaged, please replace 3061.
4. The machine is in test state when it is turned on
1. Please check whether NE556 is damaged. If it is damaged, please replace NE556.
2. Please check whether the start button is damaged. If it is damaged, please replace the start button.
5. Instrument startup, reset failure
1. Please check whether NE556 is damaged.
2. Please check whether the start and reset buttons are damaged. If they are damaged, please replace the start and reset buttons.
6. After turning on the machine and pressing the start button, the measuring light is on, but there is no voltage indication
1. Please check whether the high-voltage transformer is damaged. If it is damaged, please replace it.
2. Please check whether the voltmeter head is damaged. If it is damaged, please replace the voltmeter head.
7. No indication of leakage current
1. Please check whether the ammeter head is damaged. If it is damaged, please replace it.
2. Please check whether the 220Ω/5W resistor is damaged. If it is damaged, please replace it.
3. Please check whether LM324 is damaged. If it is damaged, please replace LM324.
Previous article:Transformer ratio tester working principle_Transformer ratio tester parameters
Next article:The difference between clamp ground resistance tester and digital ground resistance tester
- Popular Resources
- Popular amplifiers
- ICCV2023 Paper Summary: Self-, Semi-, and Unsupervised Learning
- ICCV2023 Paper Summary: Fairness, Privacy, Ethics, Social-good, Transparency, Accountability in Vision
- Nuclear Magnetic Resonance (NMR) Spectroscopy for the In Situ Measurement of Vapor−Liquid Equilibria
- Coded Signals for High Frequency Ultrasound Imaging
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
- In what situations are non-contact temperature sensors widely used?
- How non-contact temperature sensors measure internal temperature
- LED chemical incompatibility test to see which chemicals LEDs can be used with
- Application of ARM9 hardware coprocessor on WinCE embedded motherboard
- What are the key points for selecting rotor flowmeter?
- LM317 high power charger circuit
- A brief analysis of Embest's application and development of embedded medical devices
- Single-phase RC protection circuit
- stm32 PVD programmable voltage monitor
- Introduction and measurement of edge trigger and level trigger of 51 single chip microcomputer
- Improved design of Linux system software shell protection technology
- What to do if the ABB robot protection device stops
- Detailed explanation of intelligent car body perception system
- How to solve the problem that the servo drive is not enabled
- Why does the servo drive not power on?
- What point should I connect to when the servo is turned on?
- How to turn on the internal enable of Panasonic servo drive?
- What is the rigidity setting of Panasonic servo drive?
- How to change the inertia ratio of Panasonic servo drive
- What is the inertia ratio of the servo motor?
- Is it better for the motor to have a large or small moment of inertia?
- What is the difference between low inertia and high inertia of servo motors?
- Why can't the WiFi of every household in the densely packed urban communities be shared like 5G base stations?
- Teach you how to apply for a free trial of the new generation of oscilloscope 40 functional software
- Evaluation Weekly Report 20211108: Playing with domestic FPGA is full of pitfalls, but you can also have fun + Topmicro smart display module, etc.
- The msp430 microcontroller sends floating point numbers to the host computer through the serial port
- Talking about software architecture from a historical perspective - DOS era (Part 2)
- Bicycle modification series: Hot Wheels
- Participate in TE's prize-giving event and solve the problems of smart meters!
- WiFi remote control coffee shop equipment
- Tencent employees were fired for working less than 8 hours a day. What kind of work system is reasonable?
- Oscilloscope test inductor saturation current and inductance value