The oscilloscope is a commonly used electronic measuring instrument, which is widely used in many industries. When we use the oscilloscope, its measurement accuracy is very important for users. In fact, when using the oscilloscope, using some simple methods can greatly improve the measurement accuracy of the oscilloscope.
bandwidth
To accurately measure frequency response and fast rising edges, the oscilloscope and probe must have sufficient bandwidth. A good rule of thumb is that the bandwidth of the oscilloscope and probe (probes also have bandwidth limitations) should be 3 to 5 times the highest frequency of the measured signal. -3dB bandwidth attenuation will introduce a 30% amplitude measurement error, so the wider the bandwidth of the oscilloscope and probe, the better.
compatibility
It is very common to see people using an oscilloscope from Company X with a probe from Company Y. The reality is that oscilloscopes and probes are not always interchangeable or compatible. It is best practice to use oscilloscopes and probes from the same company to eliminate any potential conflict issues.
calibration
One of the most overlooked steps when taking measurements with an oscilloscope is calibration. Calibration is a simple and easy way to ensure that each measurement you make is from scratch and not affected by the last measurement. Manual calibration should be performed before starting a measurement, and if the oscilloscope has a self-calibration function, you should run this function before taking a measurement. Under-compensated or over-compensated probes can introduce significant errors in the measurement of amplitude, rise time, and waveform distortion of the measured signal.
Probe
The best measurement results are always obtained by choosing the right probe for the specific measurement task. Usually, for general-purpose measurements, a 10:1 probe is sufficient; but for low-amplitude signal measurements, you may want to consider using a 1:1 probe. When making high-speed measurements, probe capacitance should be considered. Probes with large capacitance values will slow down the rising and falling edges, and even cause some devices (such as high-speed op amps) to oscillate when probing the input or output of these devices. Therefore, another consideration for measuring high-speed circuits is to use active FET probes. Active probes usually have low contact capacitance (usually a few pF) and very high impedance, so active probes present very little loading to any node being measured.
Probe grounding
The biggest mistake you can make when making high-speed measurements is to use the probe ground clip, which can cause problems with your oscilloscope probe. Using the ground clip is like adding a series inductor to the ground path. This series inductance, combined with the probe capacitance, can introduce ringing and overshoot. The best way to make a ground connection is to use the ground shield inside the probe, but you have to disassemble the probe. However, this is not difficult to do. First, loosen the plastic cover from the probe tip and peel it off the probe. The metal ground shield of the probe is now exposed. Then, remove the ground clip from the probe and you are done. Now that you can make a measurement, simply measure a node and connect the metal ground shield of the probe to the nearest ground point of the circuit being measured. This method eliminates any series inductance and almost eliminates all ringing and overshoot. If you can't find the closest ground point, wrap a section of bus wire around the metal ground shield of the oscilloscope probe a few times and then connect it to ground. You can use almost any small metal object to make a ground connection. I have used screwdrivers, paper clips, and tweezers. My favorite tool is tweezers because the tips of the tweezers can poke into ground planes and reach into densely populated areas of the PCB.
Differential Measurement
When making a time difference or propagation delay measurement, make sure you are using two probes of the same length. The propagation delay of a cable is about 1.5ns/ft. Cable length differences can get you in trouble. For example, using a 3-foot and a 6-foot cable oscilloscope probe to measure propagation delay, the cable length difference can cause an error of about 4.5 nanoseconds (ns), which is a considerable error when trying to resolve measurements in 1ns units.
Although each of these tips and tricks may seem insignificant individually, together they can significantly improve the accuracy of your measurements. Even if you only use a few of these techniques in your measurements, they will ensure that you get fast and reliable results every time you enter the lab.
Previous article:Working process of the oscilloscope's fluorescent screen
Next article:The consequences of improper use of dual-trace oscilloscopes
- Popular Resources
- Popular amplifiers
- New IsoVu™ Isolated Current Probes: Bringing a New Dimension to Current Measurements
- Modern manufacturing strategies drive continuous improvement in ICT online testing
- Methods for Correlation of Contact and Non-Contact Measurements
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- Intel promotes AI with multi-dimensional efforts in technology, application, and ecology
- ChinaJoy Qualcomm Snapdragon Theme Pavilion takes you to experience the new changes in digital entertainment in the 5G era
- Infineon's latest generation IGBT technology platform enables precise control of speed and position
- Two test methods for LED lighting life
- Don't Let Lightning Induced Surges Scare You
- Application of brushless motor controller ML4425/4426
- Easy identification of LED power supply quality
- World's first integrated photovoltaic solar system completed in Israel
- Sliding window mean filter for avr microcontroller AD conversion
- What does call mean in the detailed explanation of ABB robot programming instructions?
- STMicroelectronics discloses its 2027-2028 financial model and path to achieve its 2030 goals
- 2024 China Automotive Charging and Battery Swapping Ecosystem Conference held in Taiyuan
- State-owned enterprises team up to invest in solid-state battery giant
- The evolution of electronic and electrical architecture is accelerating
- The first! National Automotive Chip Quality Inspection Center established
- BYD releases self-developed automotive chip using 4nm process, with a running score of up to 1.15 million
- GEODNET launches GEO-PULSE, a car GPS navigation device
- Should Chinese car companies develop their own high-computing chips?
- Infineon and Siemens combine embedded automotive software platform with microcontrollers to provide the necessary functions for next-generation SDVs
- Continental launches invisible biometric sensor display to monitor passengers' vital signs
- The concepts and differences between TTL level, CMOS level and RS232 communication level
- How to trigger ADC of C2000 series TMS320F28335?
- Measured broadband output noise of 0.8uVrms ultra-low noise regulator LT3045
- [RVB2601 Creative Application Development] Zero-based Part 2: Play with RVB2601 (based on RISC-V MCU) HelloWorld
- [Review of Arteli Development Board AT32F421] + Final Chapter: Connect the dots and solve projects
- 42 lights version of love (heart-shaped) water lamp, the romance of science and engineering men, the magic tool for chasing girls, directly on the information
- It's a power supply problem. Can anyone help me solve it?
- Principle of MCU key debounce
- Solutions to common IAR errors
- Some information about microcontroller controlling buzzer to play music