Light is an electromagnetic wave. The wavelength of 100nm to 400nm is called ultraviolet light. The light between 400nm and 780nm can be observed by the human eye, and light above 780nm is called infrared light. People can only see colors because light is reflected by objects when it is irradiated on them. The reason why green plants are green is that they absorb the red spectrum in the light. The principle of the microplate reader is to detect the absorbance value of the object under a specific wavelength.
Detection unit:
The energy difference before and after the light passes through the object is the energy absorbed by the object. At a specific wavelength, the concentration of the same object under test is quantitatively related to the absorbed energy.
The detection unit is expressed by OD value. OD is the abbreviation of optical density, which indicates the optical density absorbed by the object under test. OD = log (1/trans), where trans is the transmittance value of the object under test. According to the Bouger-amber T-beer rule, the OD value and light intensity have the following relationship:
E=OD=logΙ0/Ι, where E represents the absorbed optical density, Ι0 is the light intensity before the test object, and Ι is the light intensity from the test object.
The OD value is calculated by the following formula:
E=OD=C×D×E
C is the concentration of the test object,
D is the thickness of the test object, and
E is the molar factor
Each substance has its own specific wavelength when measured at a specific wavelength, at which the substance can absorb the most light energy. If other wavelength bands are selected, the test results will be inaccurate. Therefore, when measuring the test object, we choose a specific wavelength for detection, which is called the measurement wavelength
.
However, each substance still has a certain nonspecific absorption of light energy. In order to eliminate this nonspecific absorption, we select a reference wavelength to eliminate this inaccuracy. At the reference wavelength, the absorption of the test object light is the smallest. The difference between the absorbance values of the detection wavelength and the reference wavelength can eliminate nonspecific absorption.
Anthos ELISA instrument test value calculation
The detector in the instrument receives the light energy transmitted through the object to be tested and converts it into a binary digital signal with a maximum value of 4095. The instrument defines the light transmittance value without light source as 0% and the light transmittance value without the object to be tested as 100%. In actual testing, the light transmittance value of the object to be tested is between 0% and 100%. The calculation of transmittance value
is as follows:
T = (Meas-Min) / (Max-Min)
where T is the transmittance value, Meas is the binary value of the detection, Min is the binary value detected in the case of 0%, and Max is the binary value detected in the case of 100%. For example:
MaX = 3600 Mn = 20 Meas = 30
T = (30-20) / 3600-20) = 0.0028
OD = 1og (1/T) = 1og (1/0.0028) = 2.552
Anthos microplate reader centering
The instrument will automatically center the microplate well. The centering is to eliminate the uneven thickness caused by the convexity and concavity at the bottom of the microplate well, which brings inaccurate detection. When testing each microplate reader, the instrument actually measures 35 points, and the average of the 5 middle points is selected as the OD value of this well.
Reference channel of light source
The reference channel is used to calibrate the effects caused by unstable voltage or bulb wear.
Uses and other tips for the ELISA reader
It is used for the determination of ELISA reagents and is widely used in various laboratories, including clinical laboratories.
Quality control
Quality control is an important factor in reagent testing. Please perform quality control according to the requirements of the reagent instructions.
Blank correction
Some test kits have blank wells set to air in their instructions, while most other blank wells are set with reagents. Please follow the instructions of the kit.
Interpretation of test results
Since there are quite a few factors that affect the test results, such as different ELISA plates and the volume of test reagents, which will cause different OD values, only the test results of reagents reacted with the same ELISA plate can be compared and analyzed. Please follow the instructions of the kit for clinical interpretation of the results.
Keywords:ELISA
Reference address:Detection principle of ELISA
Detection unit:
The energy difference before and after the light passes through the object is the energy absorbed by the object. At a specific wavelength, the concentration of the same object under test is quantitatively related to the absorbed energy.
The detection unit is expressed by OD value. OD is the abbreviation of optical density, which indicates the optical density absorbed by the object under test. OD = log (1/trans), where trans is the transmittance value of the object under test. According to the Bouger-amber T-beer rule, the OD value and light intensity have the following relationship:
E=OD=logΙ0/Ι, where E represents the absorbed optical density, Ι0 is the light intensity before the test object, and Ι is the light intensity from the test object.
The OD value is calculated by the following formula:
E=OD=C×D×E
C is the concentration of the test object,
D is the thickness of the test object, and
E is the molar factor
Each substance has its own specific wavelength when measured at a specific wavelength, at which the substance can absorb the most light energy. If other wavelength bands are selected, the test results will be inaccurate. Therefore, when measuring the test object, we choose a specific wavelength for detection, which is called the measurement wavelength
.
However, each substance still has a certain nonspecific absorption of light energy. In order to eliminate this nonspecific absorption, we select a reference wavelength to eliminate this inaccuracy. At the reference wavelength, the absorption of the test object light is the smallest. The difference between the absorbance values of the detection wavelength and the reference wavelength can eliminate nonspecific absorption.
Anthos ELISA instrument test value calculation
The detector in the instrument receives the light energy transmitted through the object to be tested and converts it into a binary digital signal with a maximum value of 4095. The instrument defines the light transmittance value without light source as 0% and the light transmittance value without the object to be tested as 100%. In actual testing, the light transmittance value of the object to be tested is between 0% and 100%. The calculation of transmittance value
is as follows:
T = (Meas-Min) / (Max-Min)
where T is the transmittance value, Meas is the binary value of the detection, Min is the binary value detected in the case of 0%, and Max is the binary value detected in the case of 100%. For example:
MaX = 3600 Mn = 20 Meas = 30
T = (30-20) / 3600-20) = 0.0028
OD = 1og (1/T) = 1og (1/0.0028) = 2.552
Anthos microplate reader centering
The instrument will automatically center the microplate well. The centering is to eliminate the uneven thickness caused by the convexity and concavity at the bottom of the microplate well, which brings inaccurate detection. When testing each microplate reader, the instrument actually measures 35 points, and the average of the 5 middle points is selected as the OD value of this well.
Reference channel of light source
The reference channel is used to calibrate the effects caused by unstable voltage or bulb wear.
Uses and other tips for the ELISA reader
It is used for the determination of ELISA reagents and is widely used in various laboratories, including clinical laboratories.
Quality control
Quality control is an important factor in reagent testing. Please perform quality control according to the requirements of the reagent instructions.
Blank correction
Some test kits have blank wells set to air in their instructions, while most other blank wells are set with reagents. Please follow the instructions of the kit.
Interpretation of test results
Since there are quite a few factors that affect the test results, such as different ELISA plates and the volume of test reagents, which will cause different OD values, only the test results of reagents reacted with the same ELISA plate can be compared and analyzed. Please follow the instructions of the kit for clinical interpretation of the results.
Previous article:Theory of Eddy Current Coating Thickness Gauge
Next article:Features of infrared thermometer
Recommended Content
Latest Test Measurement Articles
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
MoreSelected Circuit Diagrams
MorePopular Articles
- Innolux's intelligent steer-by-wire solution makes cars smarter and safer
- 8051 MCU - Parity Check
- How to efficiently balance the sensitivity of tactile sensing interfaces
- What should I do if the servo motor shakes? What causes the servo motor to shake quickly?
- 【Brushless Motor】Analysis of three-phase BLDC motor and sharing of two popular development boards
- Midea Industrial Technology's subsidiaries Clou Electronics and Hekang New Energy jointly appeared at the Munich Battery Energy Storage Exhibition and Solar Energy Exhibition
- Guoxin Sichen | Application of ferroelectric memory PB85RS2MC in power battery management, with a capacity of 2M
- Analysis of common faults of frequency converter
- In a head-on competition with Qualcomm, what kind of cockpit products has Intel come up with?
- Dalian Rongke's all-vanadium liquid flow battery energy storage equipment industrialization project has entered the sprint stage before production
MoreDaily News
- Allegro MicroSystems Introduces Advanced Magnetic and Inductive Position Sensing Solutions at Electronica 2024
- Car key in the left hand, liveness detection radar in the right hand, UWB is imperative for cars!
- After a decade of rapid development, domestic CIS has entered the market
- Aegis Dagger Battery + Thor EM-i Super Hybrid, Geely New Energy has thrown out two "king bombs"
- A brief discussion on functional safety - fault, error, and failure
- In the smart car 2.0 cycle, these core industry chains are facing major opportunities!
- The United States and Japan are developing new batteries. CATL faces challenges? How should China's new energy battery industry respond?
- Murata launches high-precision 6-axis inertial sensor for automobiles
- Ford patents pre-charge alarm to help save costs and respond to emergencies
- New real-time microcontroller system from Texas Instruments enables smarter processing in automotive and industrial applications
Guess you like
- Why do different simulation results appear when the parameters of this simulation routine are set the same?
- Keil strange compilation problem
- EEWORLD University Hall----Live Replay: TI's Latest Application of Millimeter-Wave Radar in the Automotive Field
- Design of temperature measurement system based on msp430 single chip microcomputer
- Single power supply single op amp second order filter plus DC bias (cannot be isolated)
- Book now to get a gift: 2019 Keysight Electronic Measurement Basics Online Lecture Series
- Thank you + my parents
- Useful Information | A technical expert shares 15 tips on power MOSFET [Collect it]
- New Industry, New Smart Manufacturing——STMicroelectronics Industrial Tour 2019 Chengdu invites you to participate!
- Apply for a trial of a new Tektronix oscilloscope and win a custom T-shirt!