Oswald gas analyzer operation method

Publisher:会哭的蓝精灵Latest update time:2015-11-16 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere
Oswald gas analyzer operation method  

1. Washing and adjustment:

Wash all glass parts of the instrument, apply vaseline to the ground piston, and inject absorbent into each aspirator bulb.  

2. Air washing  

The purpose is to flush the original air in the instrument with the gas sample.  

3. Sampling  

If the liquid level keeps rising, it indicates that there is a leak. Check all the joints and the ground piston, and resample after plugging the leak. If the liquid level stops at a certain position after rising slightly and no longer rises, it proves that there is no leak and you can start the measurement.  

4. Determination  

The content of CO2 and O2 can be calculated as follows: CO2 or O2 (%) = × 100   

Since the volume of the gas cylinder is 100ml, the difference between the gas cylinder readings before and after the measurement is the percentage content of the measured gas, and there is no need for calculation.

Reference address:Oswald gas analyzer operation method

Previous article:Characteristics of Transformer Ratio Tester
Next article:In-depth interpretation of the development history of infrared thermal imagers

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号