Turbidimeter measurement range and system composition description

Publisher:SereneMeadow7Latest update time:2015-11-12 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere
  Main accuracy indicators and measurement range of turbidity meter 
  Measuring range 0-20, 20-200, 200-1000, 1000-2000, 2000-4000 NTU
  Resolution 0.01, 0.1 and 1 NTU Accuracy 5% FS
  Working environment temperature 0-35℃
  Power supply voltage 220±20V, 50HZ or 8 AA batteries
  System composition of turbidity meter
  The optical system of the turbidity meter consists of a tungsten lamp, a 90° detector for monitoring scattered light, and a transmitted light detector. The instrument microprocessor can calculate the ratio of the signals from the 90° detector and the transmitted light detector. This ratio calculation technology can correct for interference caused by color and/or light-absorbing substances (such as activated carbon) and compensate for the effects of light intensity fluctuations, providing long-term calibration stability. The design of the optical system can also reduce drift light and improve the accuracy of the test.
 
Reference address:Turbidimeter measurement range and system composition description

Previous article:Main features and product parameters of transformer loss parameter tester
Next article:Briefly describe the working principle of dynamic balancing tester

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号