Pump: The seal of the pump is the most easily worn part. Damage to the seal can cause many system failures. Some people recommend replacing the seal every three months. But this cannot be generalized. It depends on the material quality of the seal, the pressure used, maintenance, and the buffer. The following measures can extend the life of the seal:
1. Clean the buffer liquid in the pump every day to prevent salt deposition. The pump should be immersed in a solution without buffer or in an organic solvent.
2. Use HPLC grade reagents;
3 Use sintered stainless steel sinkers. Be careful to prevent excessive pressure from damaging the plunger rod or burning the motor due to pump blockage.
Injector: After shutdown, use solvent to rinse out the residual samples and buffer salts in the injector to prevent inorganic salt deposition and sample particles from causing wear or blockage of the valve rotor surface. It is strictly forbidden to use a sharp needle like that of a gas chromatograph for injection.
Chromatographic columns: There are several measures to prevent column performance degradation:
1. The chemical corrosion of the solvent cannot be too strong;
2. Prevent particles from settling at the column head;
3. A pressure limiter should be installed on the pump to prevent excessive pressure and large impact;
4 When the mobile phase pH is > 7, use a large particle size of the same filler as a pre-column;
5. Add a sintered stainless steel filter to the column head and a guard column if necessary;
Detector : Keep the detector clean and flush it together with the chromatographic column after each use. It is recommended to use strong solvent to reverse flush the detection pool from time to time (disassemble the column). Use degassed mobile phase to prevent air bubbles from getting stuck in the pool. The detector lamp has a certain life span, so do not turn on the lamp when not in use.
Operation limit: Each pump has an upper pressure limit. When the system reaches the preset value, the pump will automatically stop, which can protect the column and other hardware. The upper pressure limit is set below 20MPa, and the pump gasket and injection valve are durable, but setting the upper pressure limit too low will extend the analysis time. It is advisable to set the upper pressure limit at 15-20MPa according to the test. The pressure of a new column is relatively low, and the pressure rises as the use time increases, so the set upper pressure limit should also change. In gradient elution, the pressure change caused by the change of mobile phase should be considered to avoid stopping the pump midway due to too low upper pressure limit. Systems with microcomputers also need to set the lower pressure limit, generally 0.5-1MPa, and the pump will stop automatically when the mobile phase in the reservoir is drained or leaks seriously. Sometimes the lower pressure limit is set to 0, which will also stop the pump automatically.
Previous article:A brief discussion on the operating skills of gas chromatograph
Next article:The Structure and Measurement Method of Network Analyzer
- Popular Resources
- Popular amplifiers
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
- In what situations are non-contact temperature sensors widely used?
- LED chemical incompatibility test to see which chemicals LEDs can be used with
- Application of ARM9 hardware coprocessor on WinCE embedded motherboard
- What are the key points for selecting rotor flowmeter?
- LM317 high power charger circuit
- A brief analysis of Embest's application and development of embedded medical devices
- Single-phase RC protection circuit
- stm32 PVD programmable voltage monitor
- Introduction and measurement of edge trigger and level trigger of 51 single chip microcomputer
- Improved design of Linux system software shell protection technology
- What to do if the ABB robot protection device stops
- Huawei's Strategic Department Director Gai Gang: The cumulative installed base of open source Euler operating system exceeds 10 million sets
- Download from the Internet--ARM Getting Started Notes
- Learn ARM development(22)
- Learn ARM development(21)
- Learn ARM development(20)
- Learn ARM development(19)
- Learn ARM development(14)
- Learn ARM development(15)
- Analysis of the application of several common contact parts in high-voltage connectors of new energy vehicles
- Wiring harness durability test and contact voltage drop test method
- Xilinx FPGA Development Tools
- The universal tool for impedance measurement
- Common data structures in RT-Thread system
- Advice for beginners of DLP4500 light modulation.
- RC parallel circuit bandwidth problem - 3db bandwidth position problem, please help
- When I posted a message just now, as soon as I selected the invitees and clicked the Save button, an error message appeared: Discuz!
- EEWORLD University - In-depth technical training on battery management
- The firmware flashing tool is perfectly adapted to esp32-c3
- Ask about the selection of bidirectional thyristor
- 2-way ADC0808 measurement digital tube display assembly program