Causes: The formation of bubbling gas in liquid has two ways: inhalation from the outside and conversion of dissolved gas (air) in liquid into free bubbles. If the liquid contains large bubbles, it can cover the entire electrode when rubbing against the electrode, causing the flow signal input circuit to be open instantly, resulting in the output signal to fluctuate.
Judgment method: The simplest judgment method is to cut off the excitation circuit current of the magnetic field when encountering shaking. If the instrument still has a display and is unstable at this time, it means that it is mostly caused by the influence of bubbles. If the electrode resistance is measured with a pointer multimeter at this time, the loop resistance of the electrode can be measured to be higher than normal, but this test requires the long-term accumulated test experience and data of professionals.
Solution: Regarding the situation where the measured medium contains air, if it is judged to be caused by the installation position, such as the electromagnetic flowmeter is installed at the high point of the pipe system and retains gas or the air is inhaled from the outside to cause the flowmeter to shake, changing the installation position is the most thorough solution, as shown in Figure 1, at the lowest point of the pipeline or using a U-shaped tube installation. However, in many applications, the diameter is large or the installation position is not easy to change. It is recommended to install an air collection bag and an exhaust valve upstream of the flow meter, as shown in Figure 2. For a DN2200 electromagnetic flowmeter, the fluctuation of the display due to bubbles can reach 20-50%. After the exhaust device is installed, the measurement returns to normal.
Reference address:How to deal with the phenomenon of bubbles in the liquid measured by electromagnetic flowmeter
Judgment method: The simplest judgment method is to cut off the excitation circuit current of the magnetic field when encountering shaking. If the instrument still has a display and is unstable at this time, it means that it is mostly caused by the influence of bubbles. If the electrode resistance is measured with a pointer multimeter at this time, the loop resistance of the electrode can be measured to be higher than normal, but this test requires the long-term accumulated test experience and data of professionals.
Solution: Regarding the situation where the measured medium contains air, if it is judged to be caused by the installation position, such as the electromagnetic flowmeter is installed at the high point of the pipe system and retains gas or the air is inhaled from the outside to cause the flowmeter to shake, changing the installation position is the most thorough solution, as shown in Figure 1, at the lowest point of the pipeline or using a U-shaped tube installation. However, in many applications, the diameter is large or the installation position is not easy to change. It is recommended to install an air collection bag and an exhaust valve upstream of the flow meter, as shown in Figure 2. For a DN2200 electromagnetic flowmeter, the fluctuation of the display due to bubbles can reach 20-50%. After the exhaust device is installed, the measurement returns to normal.
Figure 1 Changing the installation position
Figure 2 Electromagnetic flowmeter with exhaust device installed
Previous article:802.11ac Testing with NI PXI Vector Signal Transceivers
Next article:Application of controls in intelligent inspection equipment
Latest Test Measurement Articles
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
MoreSelected Circuit Diagrams
MorePopular Articles
- Innolux's intelligent steer-by-wire solution makes cars smarter and safer
- 8051 MCU - Parity Check
- How to efficiently balance the sensitivity of tactile sensing interfaces
- What should I do if the servo motor shakes? What causes the servo motor to shake quickly?
- 【Brushless Motor】Analysis of three-phase BLDC motor and sharing of two popular development boards
- Midea Industrial Technology's subsidiaries Clou Electronics and Hekang New Energy jointly appeared at the Munich Battery Energy Storage Exhibition and Solar Energy Exhibition
- Guoxin Sichen | Application of ferroelectric memory PB85RS2MC in power battery management, with a capacity of 2M
- Analysis of common faults of frequency converter
- In a head-on competition with Qualcomm, what kind of cockpit products has Intel come up with?
- Dalian Rongke's all-vanadium liquid flow battery energy storage equipment industrialization project has entered the sprint stage before production
MoreDaily News
- Allegro MicroSystems Introduces Advanced Magnetic and Inductive Position Sensing Solutions at Electronica 2024
- Car key in the left hand, liveness detection radar in the right hand, UWB is imperative for cars!
- After a decade of rapid development, domestic CIS has entered the market
- Aegis Dagger Battery + Thor EM-i Super Hybrid, Geely New Energy has thrown out two "king bombs"
- A brief discussion on functional safety - fault, error, and failure
- In the smart car 2.0 cycle, these core industry chains are facing major opportunities!
- The United States and Japan are developing new batteries. CATL faces challenges? How should China's new energy battery industry respond?
- Murata launches high-precision 6-axis inertial sensor for automobiles
- Ford patents pre-charge alarm to help save costs and respond to emergencies
- New real-time microcontroller system from Texas Instruments enables smarter processing in automotive and industrial applications
Guess you like
- Exclusive violent disassembly of active noise reduction neck-worn voice smart headphones
- Introduction to TI_DSP link command file (*.cmd)
- 【Iprober 520 current probe】Magnetic field mode test and test summary
- RT-Thread Reading Notes] RT-Thread Study Chapter 5-6 Reading Notes
- Basic knowledge of MCU
- Writing Verilog Code for AD Sampling FPGA Program
- capacitance
- The basic principles and methods of SRIO error handling
- FPGA button problem
- How to effectively choose a lithium battery protection board