Programmable 2.4GHz
Universal Counter
Highly accurate frequency measurements to 2.4GHz
Measures Frequency, Period, Duty Cycle, RPM, Frequency ratio, Difference
frequency, Time interval and Totalize
Nine measurement functions
Plots trend graph on LCD display
Compare, Relative Error and Min/Max functions
Standard RS-232 with optional GPIB interface
Eight setup screens may be saved and retrieved from memory
Software for controlling, displaying data and data logging
Operates on all line voltages from 85V to 270V without any changes
C3100
C3100
SPECIFICATIONS
Frequency Measurement
Freq A
Ranges:
DC coupled: 0 to 120MHz AC coupled: 10Hz to 120MHz
Resolution:
0.00001Hz @ 200Hz
Gate Times:
50ms to 10s (20 settings)
Accuracy:
±2PPM;
Coupling:
AC or DC
Trigger Level:
±1.8V in a scale from -99 to +99
Freq B
Ranges:
DC coupled: 0 to 120MHz
AC coupled: 10Hz to 120MHz
Resolution:
0.00001Hz @ 200Hz
Gate Times:
50ms to 10s (20 settings)
Accuracy:
± 2PPM;
Coupling:
AC or DC
Trigger Level:
±1.8V in a scale from -99 to +99
Freq C
Ranges:
100MHz to 2.4GHz
Resolution:
100Hz @ 500MHz
Gate Times:
50ms to 10s (20 settings)
Accuracy:
±2PPM + 1 count
Coupling:
AC only
Trigger Level:
Auto
Totalize
Input:
CH A
Frequency Range:
DC to 10MHz
Count Capacity:
0 to 99,999,999 counts
Resolution:
1 count
Accuracy:
±1 count
Time Interval
(A→B)
Input:
CH A and CH B
Range:
0.5µs to 200,000µs
Minimum Pulse Width:
250ns
Resolution:
1µs
Accuracy:
±1 count + timebase accuracy
Ratio
(A/B)
Inputs:
CH A and CH B
Range:
CH A: 10MHz to 150MHzCH B: 0.1MHz to 10MHz
Resolution
(0.000001)
Duty Cycle
Input:
CH A
Measurement Range:
0.01% to 99.99%
Frequency Range:
0 to 100kHz
RPM
Input:
CH A
Measurement Range:
0 to 600,000 RPM
Minimum Pulse Width:
250ns
Resolution:
0.1 RPM
Input Characteristics
Channel A & B
Frequency Range:
0 to 120MHz
Sensitivity:
25mV
Input Impedance:
1MΩ + 35µF capacitance
Attenuator:
X1, X10
Maximum Input Volts:
250V DC or AC peak
Channel C
Frequency Range:
100MHz to 2.4GHz (usable to 2.7GHz)
Sensitivity:
25mV RMS (100MHz to 2.4GHz)
Maximum Input voltage:
5V DC or AC peak
Attenuator:
None
Input Impedance:
50Ω
Reference Timebase Oscillator
Standard Frequency:
10MHz, 4.194304MHz
Frequency Stability:
±5.0 PPM Max
Aging Rate:
±1.0 PPM Max/Yr
Temperature:
-30 to +60°C
Storage Temperature Range:
-40 to +85°C
Software
The supplied software is a Windows®-based program, which
runs under Windows 95/98/ME/XP/2000. It allows the user to
set the Instrument Parameters from the PC via the RS-232
interface and provides time stamped data logging for all
functions.
General Specifications
Display:
128 × 64 pixel super twist LCD
Line Voltage:
AC input Volts: 85 to 270 V AC ±10%
Line Frequency:
48Hz to 66Hz
Power Consumption:
15W
Operating Temperature:
0 to 40°C (32 to 104°F)
Size:
3.3” H x 11.6” W × 11.0” DWeight: 4.4 lbs.
Supplied Accessories:
Manual, Line cord, BNC cable,
Software, RS-232 cable
Optional Accessory:
GPIB (installed)
Protek Test & Measurement
40 Boroline Road, Allendale, NJ 07401 Tel: 201-760-9898 Fax: 201-760-9888
Website: www.protektest.com E-Mail: sales@protektest.com
27