Antai Oscilloscope Repair teaches you how to use the oscilloscope effectively

Publisher:彩虹微笑Latest update time:2022-04-19 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

      Antai Maintenance Huagong has been engaged in the power supply industry for 5-6 years. The oscilloscope is equivalent to my right and left hands. I feel that I can't do anything without it. With its existence, I can smoothly complete many project designs and problem analysis. For me, its contribution to today is irreplaceable. For power supply engineers, once there is a problem with a product, it is necessary to capture the waveform, capture the timing, and test the accurate values ​​to help engineers analyze and deal with it. Speak with facts and look at the waveform. How to make the test data accurate and reliable is very important. Accurate numbers can help us, while distorted waveforms and values ​​can only mislead us, make us go the other way, make us lose our direction, and do a lot of useless work.


  Thinking about it carefully, although I am not so proficient in oscilloscope research, I have read a lot of articles about oscilloscopes, encountered many problems in practice, and solved many problems. I still have some experience to share with you along the way, and I hope it can be helpful to you. Please forgive me if it is not well written.


  I often see that many small companies use oscilloscopes that are too low-end, with low bandwidth and low sampling rate. They think that it is enough to capture the waveform, and they think that there is no need to buy such a good oscilloscope. They also think that the oscilloscope is simple to operate and does not have so many specifications. When they operate the oscilloscope, they do not prepare before the test and just use it. In fact, it is incorrect to do so. It may often be this incorrect operation that causes distortion of the test results and affects the analysis. Even some very experienced engineers may not notice some details. Many engineers lack understanding of oscilloscopes, and how to use oscilloscopes better still needs to be improved. The following is to correct the common problems that many engineers I have seen and share some of the knowledge I have mastered.


  1. Many engineers just pick up the probe and test it without checking whether the probe needs compensation or the oscilloscope needs calibration. Only some large companies or trained engineers will do preparation before use.

  The oscilloscope needs to be self-calibrated and the probe compensation adjusted before use. This adjustment is performed to match the probe to the input channel.

  When operating the instrument for the first time and when displaying data from multiple input channels simultaneously, it may be necessary to calibrate the data vertically and horizontally to synchronize the time base, amplitude, and position. This may be necessary, for example, when significant temperature changes (>5°) occur.

  1. Disconnect any probes or cables from the channel input connectors. Make sure the instrument is running and has warmed up for a while. From the RFile menu, select Selfalignment.

  2. On the Control tab, click Start Alignment.

  3. In the Ralignmentstate field. The results of each calibration step for each input channel are displayed in the Results tab.

The steps for adjusting the probe compensation are as follows: 1. Connect the oscilloscope probe to the channel and press the PRESET button on the front panel (in the left panel setting area). Connect the probe signal end and reference ground to the reference output on the oscilloscope panel and press Autoset. If you use the probe hook front end accessory, firmly connect the signal pin front end to the probe to ensure a correct connection. As shown in Figure 1:

 


Oscilloscope use-1

Oscilloscope use-2

 


  Figure 1 Probe compensation adjustment


2. Check the shape of the displayed waveform. The possible situation is shown in Figure 2.


Oscilloscope use-3

 


  Figure 2 Overcompensation, undercompensation and correct compensation

  Both excess and deficiency require adjustment of the probe to better test the accurate value.

3. If the waveform is incorrect, please adjust the probe, as shown in Figure 3 below, until the waveform is the correct compensation waveform above.


Oscilloscope use-4

 


  Figure 3. Probe compensation method

  The above two points seem simple, but they are often overlooked by engineers. In order to make the measurement more accurate, please be sure to check them. These two calibration functions should be available in any oscilloscope.


  2Test voltage ripple

  Many power supply engineers do not pay much attention to the ripple measurement and just test it as they please. The different ways of using the oscilloscope lead to very different test results. As shown in Figure 4 and Figure 5 below, for the same product and the same test point, the test results are very different due to the difference in test methods. Ripple is an important parameter for power supply, but it is not worth it to fail the test due to one's own operation problems and waste a lot of manpower and cost to rectify it.

  Sometimes your customers may not pay enough attention to the use of the instrument, resulting in incorrect test data. However, there is no problem with your product, and no matter how you explain it, it doesn't make sense, so that the customer thinks you are deceiving them. Therefore, the test method is very important. Paying attention to these details can save a lot of time and improve your ability.

  The values ​​tested by the oscilloscope have errors (I will not explain this here for the time being). Now many companies require the values ​​of the test waveform as the basis for judgment. In fact, the oscilloscope only tests the process of voltage change over time, mainly to capture waveforms during debugging. The specific accuracy of measuring the effective value of DC voltage is not as good as the value of a digital multimeter. The DC accuracy of the oscilloscope is also calibrated with a multimeter as a reference. However, more and more companies and engineers use the values ​​of the oscilloscope as the real values, so we can only try our best to minimize the test error.

Below is a diagram and analysis of the test ripple:


Oscilloscope use-5

Oscilloscope use-6

Oscilloscope use-7

Oscilloscope use-8

 


  Figure 4 Figure 5 The test ripple result value of Figure 4 is 3.9921V, which is much larger than Figure 5's 0.126V. However, the test value of Figure 4 is not real. Problem analysis: In fact, there is no problem with the product. It's just that the test method is problematic. Now let's point out the problem:

  The first mistake is using a long ground wire.

  The second mistake is to place the probe loop and ground lead close to the power transformer and switching components.

  The third mistake is excess inductance between the oscilloscope probe and the output capacitor.

  Due to these carelessness, a lot of high-frequency signals, the magnetic field of the transformer, and the electric field of the switch are picked up, so that the waveform captured by the oscilloscope is displayed with high-frequency noise mixed in it.

  The fourth mistake is that the range is too large.

  To accurately test ripple, you need to:

  Use bandwidth limiting to measure ripple to prevent picking up high-frequency noise that does not really exist. Set the oscilloscope bandwidth to 20M. Remove the probe "cap" and ground clip to prevent the antenna effect formed by the long ground wire. Use a near-ground wire to wrap between the probe and the ground. Rohde & Schwarz has specially provided matching short ground wires. You can consider connecting a 0.1uf and a 10uf capacitor in parallel between the signal and the ground for decoupling. The length of the capacitor's PIN also affects the test value.

  3. Due to the lack of understanding of oscilloscopes, many engineers may misoperate and damage the oscilloscope or power supply without knowing why.

  When many junior engineers use multiple probes to measure power supply, the power supply product "explodes" as soon as it is turned on, and even damages the oscilloscope. They will ask me, doesn't the oscilloscope directly connect the probe between the components to be tested? I don't think I connected it wrong, why is it like this? That is because they don't understand the connection method of the channel and ground of the oscilloscope. The multiple probes of the oscilloscope are grounded inside the oscilloscope. So when measuring the primary and secondary sides of the power supply at the same time, if one probe is connected to the ground of the primary side and the other probe is connected to the ground of the secondary side, since the ground of the internal channel of the oscilloscope is connected together, it is equivalent to short-circuiting the primary and secondary grounds of the power supply. However, there is a voltage difference between the primary and secondary grounds, so the large current after the short circuit is easy to burn the product and the probe, and may even damage the oscilloscope. When testing the primary and secondary voltages, a differential probe should be used on one side and a normal probe on the other side. Even if the same side of the line is tested, the ground wire of the probe should be a common reference point. The ground of the oscilloscope is connected through the power ground. Many companies basically add an isolation transformer in front of the oscilloscope, which is a good method. Some companies directly cut off the PIN pins of the three-phase ground of the power supply. In this way, there is no grounding. If you touch the oscilloscope case with your hand, the leakage current will increase. It is not recommended to use it in this way.

  In fact, there are more problems, such as dynamic applications, applications of calculations between probes, and matters needing attention in testing voltage values. We all know that oscilloscopes are very powerful, and there are almost no electronic engineers who do not use oscilloscopes. Therefore, when using oscilloscopes, we must think more, do more experiments, and learn more about the functions of oscilloscopes, the differences between internal option keys, and the impact of different oscilloscope parameters on measurements, which can help us better. Don't just do it casually to complete the task. Do it seriously and observe carefully, so that we can make great progress. Experience is accumulated step by step.

[1] [1]
Reference address:Antai Oscilloscope Repair teaches you how to use the oscilloscope effectively

Previous article:Introduction to the 3000 X-Series Oscilloscopes
Next article:How to deal with problems encountered in oscilloscope measurement?

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号