Hipot Tester Calibration Steps

Publisher:CrystalBreezeLatest update time:2016-05-16 Source: eefocus Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere
1. Tester output voltage:

Hipot Tester Calibration Steps 
                                            Figure 5  Output voltage calibration wiring diagram 
        Put the withstand voltage tester in the "reset" state, turn the "voltage adjustment" knob counterclockwise to the bottom, and connect the wires according to Figure 5. Press the "start" button and adjust the "voltage adjustment" button to make the reading of the high-voltage meter the value of the detection point shown in Table 3, and calculate whether the error between the indicated value and the reading of the high-voltage meter meets the technical indicators. If not, adjust the corresponding potentiometer value. 
                                                                      Table  3

Hipot Tester Calibration Steps
Note: Different output voltages should be selected with appropriate high-voltage meter measurement range. The error of the high-voltage meter should be within ±1.5%. If it is a pointer-type high-voltage meter, its measurement range should be above 1/3 of the scale. 
Leakage current and alarm current

Hipot Tester Calibration Steps

                              Figure 6 Leakage current and alarm current wiring diagram

2) Leakage current and alarm current 
      Set the withstand voltage tester to the "reset" state, turn the "voltage adjustment" knob counterclockwise to the bottom, connect according to Figure 6, and set the "leakage current range"  switch to 2mA. Select the corresponding load resistance according to Table 4.

 

                                                                                 Table 4

Hipot Tester Calibration Steps

Reference address:Hipot Tester Calibration Steps

Previous article:How to measure the maximum range with a clamp ground resistance tester
Next article:How to use and operate the withstand voltage tester

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号