1 Introduction
Pointer-type and light-spot galvanometers have large external critical resistance and internal resistance, and have large losses in the circuit. Moreover, when the current in the energized coil changes, the coil makes a damping motion, and it takes a certain amount of time to reach a stable position. The galvanometer responds slowly, so it cannot detect instantaneous current changes and instantaneous currents with small circuit losses (such as LC oscillation currents), and is not suitable for measuring instantaneous circuit currents with small circuit resistances. Usually, multimeters can only measure the effective value of AC current and the magnitude of DC current. Therefore, neither galvanometers nor multimeters can meet the needs of measuring and observing instantaneous current changes. This design uses the principle of short-circuit current amplifier to amplify the detection current by 1:1, and can be combined with auxiliary circuits to qualitatively detect the magnitude and direction of change of transient currents with the help of light-emitting diodes.
2 Principle and Implementation
2.1 Short-circuit current amplifier
With the characteristics of high conversion rate, small input base current and drift current, and small drift voltage temperature coefficient of the integrated operational amplifier circuit, the principle of short-circuit current amplifier is used to amplify the current to be detected by 1:1, widen the required range of the internal resistance of the signal, improve the sensitivity of detection, and realize the detection of instantaneous current changes. Figure 1 shows an inverting input proportional operational amplifier circuit. The input signal Vi is connected to the inverting input terminal ∑ of the integrated operational amplifier through resistor R1, and the non-inverting input terminal ∑ is grounded through resistor R2. The output voltage VO is connected back to the inverting input terminal through the feedback resistor RF, forming a deep voltage negative feedback. In practical applications, in order to ensure that the two input terminals of the operational amplifier are in a balanced working state and avoid the input bias current generating additional differential input voltage, the resistance of the inverting input terminal and the non-inverting input terminal to ground should be equal. In Figure 1, R2=R1∥RF should be made. Since the ideal op amp's L=L=0, there is no voltage drop on R2, VO=0, and since the ideal op amp's V+=V-, V-=0, Vi=Ii×R1, so the equivalent input resistance of the inverting input amplifier circuit is r1=V/Ii=Ii×R1/Ii=R1. If R1=0, the amplifier input resistance is zero. According to the value requirement of the balancing resistor, R2=R1∥RF, then R2=0, which constitutes a short-circuit current amplifier. The impedance of the current input is zero, and the output voltage VO changes linearly with the input current. As shown in Figure 1, because V∑∑=0, it is equivalent to a short circuit in the external circuit of the signal source, but it is actually not broken. The resistance between ∑ and ∑ is extremely large, and because the resistance of the ∑ point to ground reaches several megohms, the output current of the signal source can only form a loop through RF and IC, that is, VO=IORF.
2 Working principle of detection circuit
The circuit principle of the transient current detector is shown in Figure 2. The TL084 junction field effect tube input operational amplifier is selected. Each of the operational amplifiers uses a high voltage junction field effect tube and a bipolar tube on a single integrated circuit, which is compatible with better matching, and has the characteristics of high conversion rate, small input base current and input drift current, and low drift voltage temperature coefficient. The integrated operational amplifier A and R1 form a short-circuit current amplifier, and B and R2~R6, W1 form an inverting adder, which amplifies the output voltage V1 of the operational amplifier A. Among them, R3, R4 and W1 form a circuit. If the output voltage VO of the operational amplifier B is VO≠0 when the input i=0, the movable contact of the multi-turn potentiometer W1 can be moved to make VO=0. In fact, W1 is equivalent to the role of the zero adjustment knob of the pointer type galvanometer. The voltage amplification factor of the operational amplifier B is AV=-R6/R2=-40. R7~R17 are connected in series to generate 10 reference voltages. Each integrated operational amplifier is connected to a voltage comparator, and forms a level indication circuit with a resistor and a light-emitting diode. When there is an input current i, the output voltage V1 of operational amplifier A is -i?R1. This voltage is amplified by the amplifier circuit composed of B, R2~R6, and W1, and then compared with the reference voltage of the voltage comparator. The level indication circuit formed by the light-emitting diode synchronously reflects the relative size, direction and change law of the current flowing between terminals a and b. The five voltage comparators composed of operational amplifiers C~G have their inverting input terminals connected to the reference voltages 1.918 9 V, 1.465 8 V, 1.012 7 V, 0.559 6 V, and 0.106 5 V respectively, and the non-inverting input terminals are connected to the output VO of amplifier B, which is used for the comparison and display of the forward current (i.e., flowing in from terminal a and out from terminal b). The non-inverting input terminals of the other five voltage comparators are connected to the reference voltages, -0.106 5 V, -0.559 6 V, -1.012 7 V, -1.465 8 V, and -1.918 9 V, respectively, and the inverting input terminals are all connected to the output VO of the amplifier, which are used for comparison and display of negative current (i.e., flowing in from terminal b and out from terminal a).
When current i flows from terminal a, if the magnitude is 0.1 mA, the output voltage of amplifier B is VO=0.000 1 A×330 Ω×40=1.32 V, which is higher than the reference voltage of voltage comparators E, F, G. They output high level, corresponding to LED3~LED5, which emit light; when i=0.15 mA, VO=0.00 015 A×330 Ω×40=1.98 V, which is higher than the reference voltage of voltage comparators C, D, E, E, G. These comparators output high level, corresponding to LED1~LED3 emit light. The number of LEDs is proportional to the magnitude of the detection current. When the input current changes from small to large, the order of the LEDs lighting up is LED5~LED4~LED3~LED1~LED1. When current flows from terminal b, the output voltage VO of amplifier B is negative, and comparators H, j, k, 1, M, which are responsible for negative current detection, output high level in turn, making the corresponding LED emit light. The larger the current, the lower the VO, the more light-emitting tubes light up, and the lighting order is LED7~LED8~LED9~LED10~LED11. In this way, the direction and size of the detection current can be qualitatively reflected by the number and position of the ten LEDs (except the normally lit LED6) arranged in an arc. The changes of the light-emitting tubes and the current are displayed synchronously, which is very vivid and intuitive.
Since the output voltage V1 of op amp A is equal to the product of the measured current i and R2, that is: V1=-iR1. V1max=-4.8 V. If R1=330 Ω, the maximum measurable current is imax=V1max/R1=14.55 mA. The total resistance RS of resistors R7~R17 in series is 22.07 kΩ. The voltage across R12 is V12=(5 V+5 V)R12/RS=10×470/(22.07×103)V=0.212 96 V, and the reference voltages of op amps G and H are V12/2=0.106 5 V and -V12/2=-0.106 5 V respectively. The input voltage corresponding to the output voltage of 0.106 5 V is 0.106 5/40=0.002 7 V, which is greater than the input error voltage of TL084. Assume that the minimum current that can be detected is imin. Since imin×R1×AV≥V12/2, imin≥V12/(2R1 Av)=0.106 5/(330×40)A=8.06×10-6A, so the current detection range of the galvanometer is 8.06×10-6A~14.55×10-3A. The reference voltage between two adjacent op amps, such as op amps E and F, that drive and display the current in the same direction is V=10 V×R10/RS=10 V×1 kΩ/(22.07 kΩ)V=0.453 1 V. Assuming the distinguishable input current is △i, then △iR1Av=V, so △i=V/(R1Av)=0.4531/(330×40)A=3.43×10-5A, so the current that can be displayed is 3.43×10-5A. [page]
By closing the switch K2, the direction of current change and the qualitative change of current can be observed with the help of the light-emitting diode. By closing the switch K3, the magnitude of current change can be observed with a voltmeter, but due to the inherent reasons of the voltmeter, the voltmeter cannot reflect the actual magnitude of the current with a higher frequency.
2.3 Current detector production
The operational amplifiers A to M use three TL084 operational amplifier integrated circuits. Each of them contains four identical operational amplifiers. They share a common power supply and work independently. The light-emitting diode LED6 is red or yellow, and the other LEDs are green. They are all φ6 high-brightness. All resistors are 1/8 W metal film resistors with an accuracy of 1±%. C1 uses an electrolytic capacitor with a withstand voltage of 25 V, and W1 uses a multi-turn potentiometer with a resistance of 1 kΩ. The power supply uses a ±5 V dual-output regulated power supply. Except for the potentiometer W1, LED and current-limiting resistors R18 to R28, the remaining components are designed to be installed on a copper-clad board. The circuit board diagram is shown in Figure 3. All current-limiting resistors are soldered together with the LED, which can reduce the lead wires. In order to adapt to the observation habits of the pointer galvanometer, LED1 to LED11 can be installed on the panel in a fan-shaped arrangement, with LED6 in the middle, as shown in Figure 3. In order to facilitate students' observation, the external dimensions of the current detector can be appropriately larger, such as: 40 cm high, 25 cm wide, and 10 cm thick. As long as the components are correct and installed correctly, they can work normally without debugging.
3 Applications
Before use, turn on the power switch K1, adjust W1 so that the LED6 in the middle of the row lights up, and the rest of the LEDs do not light up (i.e. adjust to zero). Connect the terminals a and b to the circuit to be tested. You can then perform the demonstration experiment just like a pointer galvanometer.
Demonstration of LC electromagnetic oscillation: Select a coil with a magnetic core with large inductance and small internal resistance. The oscillation period should be large. The special self-inductance coil of the J2343 electromagnetic oscillation demonstrator is used. Its maximum inductance is greater than 500 H and its resistance is less than 50 Ω. The capacitor is preferably a 0.6μF CBB capacitor. If it is replaced by an ordinary capacitor with a withstand voltage greater than 25 V, the reverse leakage is more serious, which accelerates the loss of energy and shortens the duration of the oscillation. The power supply is 6 V. Connect and operate according to the requirements of the demonstration experimental circuit, and you can clearly observe the damping oscillation with the same period. Due to the small damping of the LC oscillation circuit, the number of oscillations can be observed more than 5 times, and the oscillation with a period of one tenth of a second can be observed, while the pointer galvanometer can generally only observe 2 periods, and it is difficult to respond to periods less than 1 s. The relationship between the oscillation period and the capacitance can be verified by replacing capacitors of different capacities.
The demonstration of electromagnetic induction phenomenon of a single wire uses a 50-80 cm soft wire, with both ends connected to the galvanometer terminals a and b respectively. The middle part of the wire is placed in the magnetic field of a horseshoe magnet to cut the magnetic lines of force, and the galvanometer shows that an induced current is generated. The relationship between the directions of the magnetic field, wire movement, and induced current can be clearly verified, that is, the right-hand rule, which solves the experimental demonstration problem of a single wire cutting the magnetic lines of force.
Demonstration of the generator principle Connect the output end of the single-phase AC generator model to the input ends a and b of the galvanometer, so that the rotor of the generator slowly rotates from the neutral plane position. After one rotation, the LED displays the magnitude and direction of the current changes in one cycle. The faster the rotation speed is gradually increased, the faster the light-emitting tubes on both sides alternately display, and the more light-emitting tubes light up, but they are always synchronized with the rotation. When the rotation speed reaches a certain level, the alternating flashes of the light-emitting tubes on both sides gradually speed up to the point where the direction changes cannot be distinguished, and they almost always emit light. This just explains why 50 Hz AC passes through the light bulb without seeing the light flickering. This is also the observation effect that the pointer galvanometer cannot achieve.
4 Conclusion
This instantaneous current detector overcomes the inherent disadvantages of pointer-type and cursor-type galvanometers, which have large losses and slow response speeds in the circuit. It can detect instantaneous current changes and is suitable for detecting instantaneous currents in the circuit to be tested that require small losses and small circuit resistance. As a low-cost measuring instrument for related physical and electrical measurements, it can be used for demonstration experiments such as LC electromagnetic oscillation, single-wire electromagnetic induction, and generator principles.
Previous article:Nonlinear Measurements Based on PNA-X NVNA Network Analyzer
Next article:Design of MIL-STD-1553A/B Data Bus Automatic Tester
- Popular Resources
- Popular amplifiers
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
- Innolux's intelligent steer-by-wire solution makes cars smarter and safer
- 8051 MCU - Parity Check
- How to efficiently balance the sensitivity of tactile sensing interfaces
- What should I do if the servo motor shakes? What causes the servo motor to shake quickly?
- 【Brushless Motor】Analysis of three-phase BLDC motor and sharing of two popular development boards
- Midea Industrial Technology's subsidiaries Clou Electronics and Hekang New Energy jointly appeared at the Munich Battery Energy Storage Exhibition and Solar Energy Exhibition
- Guoxin Sichen | Application of ferroelectric memory PB85RS2MC in power battery management, with a capacity of 2M
- Analysis of common faults of frequency converter
- In a head-on competition with Qualcomm, what kind of cockpit products has Intel come up with?
- Dalian Rongke's all-vanadium liquid flow battery energy storage equipment industrialization project has entered the sprint stage before production
- Allegro MicroSystems Introduces Advanced Magnetic and Inductive Position Sensing Solutions at Electronica 2024
- Car key in the left hand, liveness detection radar in the right hand, UWB is imperative for cars!
- After a decade of rapid development, domestic CIS has entered the market
- Aegis Dagger Battery + Thor EM-i Super Hybrid, Geely New Energy has thrown out two "king bombs"
- A brief discussion on functional safety - fault, error, and failure
- In the smart car 2.0 cycle, these core industry chains are facing major opportunities!
- The United States and Japan are developing new batteries. CATL faces challenges? How should China's new energy battery industry respond?
- Murata launches high-precision 6-axis inertial sensor for automobiles
- Ford patents pre-charge alarm to help save costs and respond to emergencies
- New real-time microcontroller system from Texas Instruments enables smarter processing in automotive and industrial applications
- Goodbye 2019 ~ Hello 2020 - Looking back at 2019 and planning for 2020
- 【Home Smart Dashboard】Project Submission
- FPGA design ideas and techniques: data interface synchronization
- Together: About the GD32L233 development board unpacking test
- It is said that uC/OS system is free. Is it true?
- Last two days! Free review - Pingtouge scenario-based Bluetooth Mesh gateway development kit
- EEWORLD University Hall----Research and development of high-speed motors
- How to choose between different chips in TI's Sub-1GHz product line?
- Why does C6000 need Cache?
- What are the benefits of using an op amp to drive a MOS tube? Who knows? Tell me