Test system static characteristics calibration

  • 2013-09-22
  • 291.04KB
  • Points it Requires : 2

    Experimental purpose 1.1 Master the principle of pressure sensor 1.2 Master the composition of pressure measurement system 1.3 Master the general method of static calibration experiment and static calibration data processing of pressure sensor 2. Experimental equipment This experimental system consists of piston pressure gauge, silicon piezoresistive pressure sensor, signal conditioning circuit, 5.5-digit digital voltmeter, DC regulated power supply and sampling resistor. Figure 1-1 The experimental system block diagram is as follows