A brief introduction to the syringe sliding leakage tester

Publisher:书香门第Latest update time:2022-07-07 Source: elecfans Reading articles on mobile phones Scan QR code
Read articles on your mobile phone anytime, anywhere

Introduction: The cartridge bottle is also known as the borosilicate glass sleeve for pen-type syringes. The front of the bottle is equipped with an injection needle protected by a rubber seal, the bottle mouth is sealed with a rubber stopper and aluminum cap, and the tail is sealed with a rubber piston.


The cartridge leakage performance and sliding performance tester is designed and manufactured in full compliance with relevant standards such as YBB00152004-2015 "Chlorobutyl rubber piston and gasket for pen syringes", YBB00162004-2015 "Brominated butyl rubber piston and gasket for pen syringes", and YBB00112004-2015 "Prefilled syringe assemblies (with injection needles)".


Leakage test: Take 10 pen-type syringe sleeves with inner surfaces treated with silicone oil, fill them with water, install the test piston and gasket, add aluminum caps, and seal them. Install one pen-type syringe sleeve in the pen-type syringe sleeve holder (with a flat surface in contact with the piston), apply a force of 60N±3N to the piston, and keep it for 1 minute. After the test, check the leakage of the piston and gasket. The same operation is performed on the remaining pen-type syringe sleeves. There should be no liquid leakage in the piston and gasket.


Piston sliding test: Take 10 pen-type syringe barrels with inner surfaces treated with silicone oil, fill them with water, install the test pistons and gaskets, add aluminum caps, and seal them.


Install a pen syringe sleeve into the pen syringe sleeve holder (with a flat surface in contact with the piston, the holder must ensure that it can pierce the gasket installed on the pen syringe sleeve. The injection needle is a hypodermic needle with a blade at both ends and an outer diameter of 0.4mm.


When the piston moves, the water in the sleeve should be able to be discharged through the needle), move the piston downward at a speed of 40mm/min±3mm/min, pause for 5 seconds when it moves half the distance in the pen-type syringe sleeve, and then press the piston completely to the bottom. The remaining pen-type syringe sleeves are operated in the same way. Record the first force to start the piston in each test, and take the average of 10 * values ​​as the "starting force", which should not exceed 30N; record the force required to keep it moving after starting in each test, and take the average of 10 values ​​as the "continuous driving force", which should not exceed 15N; record the first force to push the piston again after pausing for 5 seconds in each test, and take the average of 10 * values ​​as the "restarting force", which should not exceed 30N; and check that there is no "tremor" (referring to the irregular movement of the piston) during continuous movement.

Reference address:A brief introduction to the syringe sliding leakage tester

Previous article:Introduction to the leading configuration of cable fault test instruments in China
Next article:Working principle and characteristics of moisture barrier microbial penetration tester

Latest Test Measurement Articles
Change More Related Popular Components

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号