The rotor flowmeter (float flowmeter) is a variable area flowmeter. The liquid flows in the flowmeter pipeline, causing the float in the pipeline to move upward. When the flow rate and flow rate are constant, the area between the float and the pipeline reaches a certain value. This area is proportional to the height of the float and is also proportional to the flow rate. The pointer of the pointer-type rotor flowmeter is displayed on the dial according to the height of the float through the rotation of the internal parts. The electronic head displays the intelligent calculation display of the circuit board.
1. The rotor flowmeter has large measurement errors:
1. Since the gas medium is greatly affected by temperature and pressure, it is recommended to use temperature and pressure compensation to obtain the true flow rate.
2. Due to long-term use and pipeline vibration and other factors, the floating flowmeter sensor magnet, pointer, counterweight, rotating magnet and other active parts are loose, resulting in large errors. Solution: You can first verify it by pushing the pointer by hand. First, press the pointer at the RP position to see if the output is 4mA and the flow display is 0%, and then verify it according to the scale in turn. If it is found to be inconsistent, the position of the component can be adjusted. Generally, professionals are required to adjust it, otherwise the position will be lost and it needs to be returned to the manufacturer for correction.
3. Installation does not meet the requirements: For vertical installation of rotor flowmeter, it should be kept vertical, with an inclination of no more than 20 degrees; for horizontal installation of rotor flowmeter, it should be kept horizontal, with an inclination of no more than 20 degrees; there should be no ferromagnetic objects in the 100mm space around the rotor flowmeter. The installation location should be away from the valve reducer, pump outlet, process pipeline turning port, etc. The requirement of 5D before and 250mm straight pipe section after should be maintained.
4. The large change in density of liquid medium is also a reason for large errors. Because the instrument is converted according to the density given by the user before calibration, and converted into the flow rate of water under the calibration state for calibration, if the density of the medium changes greatly, it will cause a large error in the measurement. The solution is to bring the density of the medium after the change into the formula, convert it into the error correction coefficient, and then multiply the flow measured by the flowmeter by the coefficient to convert it into the real flow.
2. Rotor flowmeter pointer jitter:
1. Severe pointer jitter: mainly due to medium pulsation, unstable air pressure or the pressure, temperature and flow of the gas operation state given by the user do not match the actual state of the rotor flowmeter, and there is a large difference causing the rotor flowmeter to over-range. 2. Slight pointer jitter: generally caused by medium fluctuations. It can be overcome by increasing damping. 3. Moderate pointer jitter: generally caused by the flow state of the medium. For gas, it is generally caused by unstable medium operating pressure. Voltage stabilization or flow stabilization devices can be used to overcome or increase the gas damping of the rotor flowmeter. 3. The pointer of the
rotor flowmeter stops at a certain position and does not move:
the main reason is that the rotor of the rotor flowmeter is stuck. Generally, when the rotor flowmeter is used, the valve is opened too quickly, causing the rotor to quickly impact the stopper upward, causing the stopper to deform and the rotor to be stuck. However, it is not ruled out that the rotor is stuck due to the non-concentricity of the rotor guide rod and the stop ring. When handling, the instrument can be removed, the deformed stopper can be removed and shaped, and check whether it is concentric with the guide rod. If it is not concentric, it can be corrected, and then the rotor can be installed, and the rotor can be pushed by hand. It can be felt that the rotor is unobstructed up and down without obstruction. In addition, when installing the rotor flowmeter, it must be installed vertically or horizontally, and it cannot be tilted, otherwise it is easy to cause the meter to be stuck and bring errors to the measurement.
Reference address:Three common faults of rotor flowmeter
1. The rotor flowmeter has large measurement errors:
1. Since the gas medium is greatly affected by temperature and pressure, it is recommended to use temperature and pressure compensation to obtain the true flow rate.
2. Due to long-term use and pipeline vibration and other factors, the floating flowmeter sensor magnet, pointer, counterweight, rotating magnet and other active parts are loose, resulting in large errors. Solution: You can first verify it by pushing the pointer by hand. First, press the pointer at the RP position to see if the output is 4mA and the flow display is 0%, and then verify it according to the scale in turn. If it is found to be inconsistent, the position of the component can be adjusted. Generally, professionals are required to adjust it, otherwise the position will be lost and it needs to be returned to the manufacturer for correction.
3. Installation does not meet the requirements: For vertical installation of rotor flowmeter, it should be kept vertical, with an inclination of no more than 20 degrees; for horizontal installation of rotor flowmeter, it should be kept horizontal, with an inclination of no more than 20 degrees; there should be no ferromagnetic objects in the 100mm space around the rotor flowmeter. The installation location should be away from the valve reducer, pump outlet, process pipeline turning port, etc. The requirement of 5D before and 250mm straight pipe section after should be maintained.
4. The large change in density of liquid medium is also a reason for large errors. Because the instrument is converted according to the density given by the user before calibration, and converted into the flow rate of water under the calibration state for calibration, if the density of the medium changes greatly, it will cause a large error in the measurement. The solution is to bring the density of the medium after the change into the formula, convert it into the error correction coefficient, and then multiply the flow measured by the flowmeter by the coefficient to convert it into the real flow.
2. Rotor flowmeter pointer jitter:
1. Severe pointer jitter: mainly due to medium pulsation, unstable air pressure or the pressure, temperature and flow of the gas operation state given by the user do not match the actual state of the rotor flowmeter, and there is a large difference causing the rotor flowmeter to over-range. 2. Slight pointer jitter: generally caused by medium fluctuations. It can be overcome by increasing damping. 3. Moderate pointer jitter: generally caused by the flow state of the medium. For gas, it is generally caused by unstable medium operating pressure. Voltage stabilization or flow stabilization devices can be used to overcome or increase the gas damping of the rotor flowmeter. 3. The pointer of the
rotor flowmeter stops at a certain position and does not move:
the main reason is that the rotor of the rotor flowmeter is stuck. Generally, when the rotor flowmeter is used, the valve is opened too quickly, causing the rotor to quickly impact the stopper upward, causing the stopper to deform and the rotor to be stuck. However, it is not ruled out that the rotor is stuck due to the non-concentricity of the rotor guide rod and the stop ring. When handling, the instrument can be removed, the deformed stopper can be removed and shaped, and check whether it is concentric with the guide rod. If it is not concentric, it can be corrected, and then the rotor can be installed, and the rotor can be pushed by hand. It can be felt that the rotor is unobstructed up and down without obstruction. In addition, when installing the rotor flowmeter, it must be installed vertically or horizontally, and it cannot be tilted, otherwise it is easy to cause the meter to be stuck and bring errors to the measurement.
Previous article:Common Problems of Ultrasonic Flow Meters
Next article:Principle and characteristics of balanced flowmeter
Recommended Content
Latest Test Measurement Articles
- Keysight Technologies Helps Samsung Electronics Successfully Validate FiRa® 2.0 Safe Distance Measurement Test Case
- From probes to power supplies, Tektronix is leading the way in comprehensive innovation in power electronics testing
- Seizing the Opportunities in the Chinese Application Market: NI's Challenges and Answers
- Tektronix Launches Breakthrough Power Measurement Tools to Accelerate Innovation as Global Electrification Accelerates
- Not all oscilloscopes are created equal: Why ADCs and low noise floor matter
- Enable TekHSI high-speed interface function to accelerate the remote transmission of waveform data
- How to measure the quality of soft start thyristor
- How to use a multimeter to judge whether a soft starter is good or bad
- What are the advantages and disadvantages of non-contact temperature sensors?
MoreSelected Circuit Diagrams
MorePopular Articles
- Innolux's intelligent steer-by-wire solution makes cars smarter and safer
- 8051 MCU - Parity Check
- How to efficiently balance the sensitivity of tactile sensing interfaces
- What should I do if the servo motor shakes? What causes the servo motor to shake quickly?
- 【Brushless Motor】Analysis of three-phase BLDC motor and sharing of two popular development boards
- Midea Industrial Technology's subsidiaries Clou Electronics and Hekang New Energy jointly appeared at the Munich Battery Energy Storage Exhibition and Solar Energy Exhibition
- Guoxin Sichen | Application of ferroelectric memory PB85RS2MC in power battery management, with a capacity of 2M
- Analysis of common faults of frequency converter
- In a head-on competition with Qualcomm, what kind of cockpit products has Intel come up with?
- Dalian Rongke's all-vanadium liquid flow battery energy storage equipment industrialization project has entered the sprint stage before production
MoreDaily News
- Allegro MicroSystems Introduces Advanced Magnetic and Inductive Position Sensing Solutions at Electronica 2024
- Car key in the left hand, liveness detection radar in the right hand, UWB is imperative for cars!
- After a decade of rapid development, domestic CIS has entered the market
- Aegis Dagger Battery + Thor EM-i Super Hybrid, Geely New Energy has thrown out two "king bombs"
- A brief discussion on functional safety - fault, error, and failure
- In the smart car 2.0 cycle, these core industry chains are facing major opportunities!
- The United States and Japan are developing new batteries. CATL faces challenges? How should China's new energy battery industry respond?
- Murata launches high-precision 6-axis inertial sensor for automobiles
- Ford patents pre-charge alarm to help save costs and respond to emergencies
- New real-time microcontroller system from Texas Instruments enables smarter processing in automotive and industrial applications
Guess you like
- EEWORLD University Hall----Live Replay: TI's low-power wireless technology for building automation sensor applications
- PSRR Measurement of LDO
- quartusii official tutorial
- [Repost] Learn these eight circuit design skills and your professional quality will be greatly improved
- [STM32WB55 Review] +ST's Attack
- MSP430 ADC acquisition filter
- Soil moisture measurement circuit diagram
- I tested the TLC2272 chip and found that the rail-to-rail output is really good.
- [CC1352P Review] BLE Program Framework
- EEWORLD University ---- Robotics