DATA SHEET
MBC150 SERIES MEDICAL 150W AC/DC
110 W convection cooled
-20 to 50 deg C full load operation
2 x MOPP Isolation
90-264 VAC input
2” x 4” x 1.3” (50.8 x 101.6 x 33.6 mm)
Fan output, 12 Vdc @ 0.5A standard
No minimum load required
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
FEATURES
APPLICATIONS
Dialysis
Monitoring
Pumps
Surgical Devices
Home Health
Portable Devices
BCD.00288_AA_December-3-2013
1
www.power-one.com
DATA SHEET
TECHNICAL DATA:
Input
PARAMETER
Input Voltage Range
Input Frequency Range
Input Surge Current
Safety Ground Leakage Current
Input Current
DESCRIPTION/CONDITION
Universal Input
47-63 Hz
230 Vac (cold start)
264 Vac 50 / 60 Hz
120 Vac @ Full load
230 Vac @ Full load
90 - 264 Vac
120 – 390 Vdc
65 A max.
< 250 μA max
1.7 A
0.85 A
Output
PARAMETER
Voltage Adjustment
Transient Response
Over Voltage Protection
Over Current Protection
Short Circuit Protection
Efficiency
Rise Time
Set Point Accuracy (Main Output)
DESCRIPTION/CONDITION
V1
Main output 50 to 100% load change, 50 Hz,
50% duty cycle, 0.1A / uSec
V1
Rated output current
Automatic recovery
>86%
< 100 mSec
± 1%
± 3%
< 10%, recovery time < 5 mSec
110 to 150% rated max
110% Typical
Ordering Information
PRODUCT FAMILY
MBC150-1005G
MBC150-1005G-2
MBC150-1T05G
MBC150-1T05G-2
MBC150-1012G
MBC150-1012G-2
MBC150-1T12G
MBC150-1T12G-2
MBC150-1015G
MBC150-1015G-2
MBC150-1T15G
MBC150-1T15G-2
MBC150-1024G
MBC150-1024G-2
MBC150-1T24G
MBC150-1T24G-2
MBC150-1048G
MBC150-1048G-2
MBC150-1T48G
MBC150-1T48G-2
Vfan (all models)
VOLTS
MAX LOAD
MAX LOAD
(1)
300 LFM
(1)
(VDC) CONVECTION
5.0
5.0
5.0
5.0
12
12
12
12
15
15
15
15
24
24
24
24
48
48
48
48
12
16.0 A
16.0 A
16.0 A
16.0 A
8.33 A
8.33 A
8.33 A
8.33 A
6.67 A
6.67 A
6.67 A
6.67 A
4.17 A
4.17 A
4.17 A
4.17 A
2.08 A
2.08 A
2.08 A
2.08 A
0.5 A
(3)
16 A
16 A
20 A
20 A
12.5 A
12.5 A
12.5 A
12.5 A
10.0 A
10.0 A
10.0 A
10.0 A
6.25 A
6.25 A
6.25 A
6.25 A
3.13 A
3.13 A
3.13 A
3.13 A
0.5 A
(3)
MINIMUM
LOAD (A)
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
RIPPLE &
NOISE
(4)
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
CONNECTOR
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
TOTAL
REGULATION
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
±
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
2.5%
20%
MBC150 Series
2
www.power-one.com
DATA SHEET
Notes:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Combined power from main output and Vfan should not exceed total power rating.
Fan output tolerance is ± 20%.
Peak current for fan output is 1A.
Ripple is 2% up to 20% load and less than 1% above 20% load. Output noise measurement is made with a 20 MHz
bandwidth using a 6” twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor.
Class 1 models have Earthing tab J4. Class 2 models (-2 suffix) have no Earthing tab.
Specifications are for nominal input voltage, 25° C and max load unless otherwise stated.
Air flow over length of supply recommended (either direction) for forced air rating.
Derate output power per chart below.
Specifications subject to change without notice.
Warranty 2 years.
General Specifications
PARAMETER
Hold Up Time
MTBF
Switching Frequency
Isolation Voltage
Weight
DESCRIPTION/CONDITION
120 Vac
230 Vac
>200 khrs
PFC converter variable 35 to 250 kHz, 90
kHz typical
Input to Output Min 5900 Vdc
150 g (0.33 lbs)
6 mSec
10 mSec
Bellcore TR-332
Resonant converter: Variable 35 to 250
kHz, 90 kHz typical
Input to Ground 2120 Vdc
Environmental
PARAMETER
Operating Temperature
Altitude
Humidity
Conducted Emissions
Radiated Emissions
Electromagnetic Susceptibility
Harmonic Current
DESCRIPTION/CONDITION
Operating
Storage
Operating 10,000 ft.
95%
EN55022, FCC part 15 Level B
EN55022, FCC part 15 Level B
EN61000-4
EN61000-3-2, Class D
-20 to +70° C. See derating charts below.
-40 to +85° C
Non-operation 40,000 ft.
Non-condensing
To be controlled in end system
2, 3, 4, 5 Level 3
Safety
PARAMETER
EN / UL / CSA
DESCRIPTION/CONDITION
60601-1 3rd Edition
BCD.00288_AA_December-3-2013
3
www.power-one.com
DATA SHEET
Figure 1 Output Power Vs. Temperature
Figure 2 Dimension Drawing (Top and Side View)
Mechanical
INPUT = J1
Pin 1: AC Line
Pin 2: Removed
Pin 3: AC Neutral
Mating Connector:
Molex: 09-50-3031
Pins: 08-50-0106
EARTHING TAB = J4
Molex: 19705-4301
Mating Connector:
Molex: 190030001
Pin 1 = V1
Pin 2 = V1
DC OUTPUT = J2
Pin 3 = RTN
Pin 4 = RTN
FAN= J3
Pin 1 = +12 V @ 0.5 A
Pin 2 = Fan return
(isolated from DC output)
Mating Connector:
Tyco: 640440-2
Mating Connector:
JST VHR-4M, Pins: SVH-41T-P1.1
AWG #20 to #16
Copyright © 2010 Power-One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherwise, the
trademarks and service marks of Power-One Inc. in the U.S. and other countries. All other product or service names are the property of their respective holders.
Power-One products are protected under numerous U.S. and foreign patents and pending applications, maskwork rights, and copyrights. Power-One reserves the
right to make changes to any products and services at any time without notice. Power-One assumes no responsibility or liability arising out of the application or use
of any information, product, or service described herein except as expressly agreed to in writing by Power-One Inc.
NUCLEAR AND MEDICAL APPLICATIONS
- Power-One products are not designed, intended for use in, or authorized for use as critical components in life
support systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president
of Power-One, Inc.
TECHNICAL REVISIONS
- The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.
MBC150 Series
4
www.power-one.com