GB20 Family
Medical
20W Single Output
Medical & Industrial Grade
FEATURES AND BENEFITS
20W Open Frame and PCB-mount Power Supply
1.6” x 3.38” x 1.0” Package
Test &
Measurement
Approved to CSA/EN/IEC/UL60601-1, 3
rd
Edition
E-cap Life of >7 Years
>1,000,000 Hours MTBF
3 Year Warranty
Meets Class B Radiated & Conducted EMI, with
Margin
Universal Input 90-264VAC
<0.1W No Load Input Power
Approved to CSA/EN/IEC/UL62368-1
Industrial
Meets Heavy Industrial and IEC60601-1-2
4
th
Edition Levels of EMC
Note:
*Consult Factory for compliance information.
MODEL SELECTION
Model Number
2
GB20S05K01
GB20S07K01
GB20S09K01
GB20S12K01
GB20S15K01
GB20S24K01
GB20S48K01
Volts
5.0V
7.5V
9.0V
12.0V
15.0V
24.0V
48.0V
Rated
Current
3.0A
2.0A
2.0A
1.5A
1.2A
0.8A
0.4A
Output
Power
15W
15W
18W
18W
18W
20W
20W
Ripple &
Noise
1
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
120mV pk-pk
240mV pk-pk
480mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Input
Class/Termination
Class I (Grounded) input,
3-pin AMP/Molex type
connector
Change “K” to “C” for
class II input
Change “K” to “P” for PCB
mount pins, class I input
Change “K” to “PCB mount
pins”, class II input
Output
Termination
4-pin AMP/Molex type
connector for “K” and “C”
versions
PCB mount pins for
“P” and “V” versions
Note:
1. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
2. Other output voltages available, consult factory.
3. All specifications are typical at 230VAC, full load, at 25°C ambient unless noted.
INPUT
Input Voltage and
Frequency
Input Current
Inrush Current
Input Fuses
Earth Leakage Current
(Input to Ground)
Earth Leakage Current
(Output to Ground)
Efficiency
Power Factor
100-240VAC, ±10%, 47-63Hz, 1Ø
115VAC: 0.5A, 230VAC: 0.2A
264VAC, cold start: will not exceed 40A peak
3.15A, 250VAC fuse in both line and neutral
<500μA@264VAC, 60Hz, NC
<1mA@264VAC, 60Hz, SFC
<100μA@264VAC, 60Hz, NC
<500μA@264VAC, 60Hz, SFC
>88%, typical
0.9, min., 230VAC, 80-100% load vector, 25°C
ambient
OUTPUT
Turn On Time
Hold-Up Time
Output Power
Output Voltage
<700ms
20ms/100VAC at full load
15W-20W continuous – See models chart for specific
voltage model ratings
See models chart
500μs
resp.time for return to w/in 0.5% of final value
for any 50% load step from 5% to 100% of rated load,
∆i/∆t<0.2A/μs
Max voltage deviation is +/-3.5%
See models chart
Transient Response
Regulation
Note:
All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
Note:
All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
GB20 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 1
GB20 Family
PROTECTION
Overtemperature
Protection
Overload Protection
Short Circuit Protection
Overvoltage Protection
Will shutdown upon an overtemperature condition,
Auto-recovery
130% to 160% of rated output current value,
Hiccup mode
Hiccup mode
120% to 150% of nominal output voltage,
Hiccup Mode
20W Single Output
Medical & Industrial Grade
SAFETY
ITE/Industrial Safety
Medical Safety
EN/IEC/UL62368-1
EN/IEC/UL60601-1, 3
rd
edition
RELIABILITY
MTBF
E-cap Life
>1,000,000 hours, full load, 110 & 220VAC input,
25°C amb., per telcordia 332 issue 6, stress method
>7 year life based on calculations at 115VAC/60Hz
& 230VAC/50Hz, ambient 25°C at 24 hrs per day,
365 days/year, 6 power up cycles per day
EMI/EMC COMPLIANCE
Conducted Emissions
Radiated Emissions
EN55032, EN55011/CISPR11 Class B, FCC Part
15.107, Class B: 6db margin type, at 115 and 230VAC
EN55032, EN55011/CISPR11 Class B, FCC Part
15.109, Class B: 3db margin type, at 115 and 230VAC
ISOLATION SPECIFICATIONS
Isolation
Isolation Resistance
Input-Output
: 4000VAC (2 MOPP)
Input-Ground : 1500VAC (1 MOPP)
Output-Ground : 1500VAC (1 MOPP)
I/P-O/P, I/P-FG, O/P-FG: TBD
Electro-Static Discharge
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact,
(ESD) Immunity on
+/- 15kV air, Criteria A
Power Ports
IEC60601-1-2, 4
th
edition, Table 4
Radiated RF EM Fields
Susceptibility
Electrical Fast
Transients (EFT)/Bursts
Surges, Line to Line
(Diff Mode) and Line to
GND (CMN Mode)
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz,
80% AM at 1kHz
IEC60601-1-2, 4
th
edition, Table 4
EN55024/IEC61000-4-4, Level 4, +/- 4.4kV,
100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4
th
edition, Table 5
EN55024/IEC61000-4-5, Level 4, +/-2kV DM,
+/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4
th
edition requirements
ENVIRONMENT
Operating Temperature
StorageTemperature
Cooling
Relative Humidity
-25 ~ +70°C, see derating curve for operation above
40°C
-40°C
~
+85°C
Convection
5% to 90%, Non-condensing
Operating: 0.003g/Hz, 1.5grms overall, 3 axes,
10 min/axis, 1-500Hz
Non-Oper.: random waveform, 3 minutes per axis, 3
axes and sine waveform, Vib. frequency/acceleration:
10-500Hz/1g, sweep rate of 1 octave/minutes,
vibration time of 10 sweeps/axes, 3 axes
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6 shocks
total
Non-Operating: Half-sine waveform, impact
acceleration of 50G, pulse duration of 6ms,
Number of shocks: 3 for each of the three axis
41 x 86 x 25mm
1.6 x 3.38 x 1.0 inch
120g
EN55022/IEC61000-4-6, 3.6V/m – Level 4, (0.15 to
Conducted Disturbances
80Mhz; and 12V/m) in ISM and amateur radio bands
between 0.15Mhz and 80Mhz, 80% AM at 1KHz
Induced by RF Fields
IEC60601-1-2, 4
th
edition, Table 5
Rated Power Frequency
Magnetic Fields
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60Hz
IEC60601-1-2, 4
th
edition, Table 4
EN55024/IECEN61000-4-11:
--100% dip for 10 ms, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20ms, 0 deg.,
Criteria A
--100% dip for 5000ms (250/300 cycles), Criteria B
--60% dip for 100ms, Criteria B
--30% dip for 500ms, Criteria A
IEC60601-1-2, 4
th
edition, Table 5
EN55011/EN61000-3-2, Class A
EN61000-3-3
Vibration
Voltage Interruptions,
Dips, Sags & Surges
Shock
Harmonic Current
Emissions
Flicker Test
Dimensions
Weight
Note:
1. The power supply is considered a component which will be installed into a final equipment.
The final equipment must be re-confirmed that it still meets EMC directives.
2. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult
factory for information regarding testing for or usage under special environments.
Note:
Same dimensions for PCB & Pin Variants.
GB20 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 2