C
Series
Switch with IEC Connector Compact Size
UL Recognized
CSA Certified
VDE Approved
C Series
Two function power entry module combining a DPST
switch and an IEC 320 power line connector. Compact
packaging allows a smaller panel cutout with choice of
snap-in or flange mounting. Available with general
purpose filtering (F models - CFS, CFE), medical
application filtering (H models - CHS, CHE) or unfiltered
(CS, CE). Terminals allow for easy connection to any
wiring scheme. Preconnected terminals from line to
switch are available (B Models - CBS, CBE) for reducing
OEM wiring time.
The medical versions (H models) are designed
specifically for health care applications. Their extremely
low leakage current allows compliance with UL544 and
UL2601 as well as CSA22.2 health care specifications for
use with both patient and non-patient care equipment.
The C series also has the adaptability to fit into existing
panel cutouts using snap-on extender options. Using the
CA430 extender, the panel cutout height is increased by
0.13”. (Extender is available pre-assembled using part
numbers with ‘K’ designation)
CFS
CS
Specifications
Hipot rating (one minute):
line-to-ground
line-to-line
Operating frequency:
Rated voltage:
Rated current:
Operating voltages:
Switch:
2250 VDC
1450 VDC
50/60 Hz
120/250 VAC
10A@120 VAC
10A @ 250 VAC
115/230 VAC
Rated for 10,000
operations at full load.
51 Amp inrush capability.
Unfiltered .187” x .032” terminals
Filtered
.25” x .032” terminals
Terminals:
Filtered Models Only:
Maximum leakage current, each line-to-ground
F models
H models
@ 120 VAC 60 Hz:
@ 250 VAC 50 Hz:
@ 120 VAC 60 Hz:
@ 250 VAC 50 Hz:
.25mA
.40mA
2µA
5µA
Electrical Schematic
F Models (CFS, CFE)
Minimum
insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current
Frequency MHz
Rating
0.05 0.15 0.5
1
5
H Models
1A
3A
6A
10A
F Models
1A
3A
6A
10A
Note 2
Note 2
10
21
24
18
11
47
48
41
33
30
10
14
13
9
46
50
50
42
H Models (CHS, CHE)
16
9
4
2
10
8
4
1
21
14
10
6
26
16
11
4
37
31
22
10
46
32
22
14
44
32
23
15
48
36
27
18
26
26
19
11
46
43
36
27
B Models (CBS, CBE)
Note 1
Note 1
C
86
Power on/off
Resistor location for reference only.
Note 1: Jumpers provided on CBS and CBE versions.
Note 2: Location of optional filter.
Line-to-line in 50 ohm circuit
Current
Frequency MHz
Rating
0.05 0.15
0.5
1
5
H Models
1A
1
6
13
29
38
3A
1
5
10
22
36
6A
1
5
14
20
31
10A
1
4
11
19
32
F Models
1A
1
3
13
28
62
3A
2
6
14
23
65
6A
2
6
14
27
46
10A
1
7
14
23
42
10
42
34
33
37
67
65
48
44
30
26
36
37
38
42
67
58
62
www.cor.com - Libertyville, IL (847) 680-7400
Series
C
Case Styles
Metric shown in italics.
Case Dimensions
Metric shown in italics.
Part
No.
CS, CBS
CS, CBS
†
A
(max)
B
(max)
.97
24.64
.97
24.64
2.53
64.26
2.53
64.26
C
(max)
1.62
41.17
1.62
41.17
1.62
41.17
1.62
41.17
D
±.006
±.152
1.06
26.92
1.06
26.92
1.12
28.5
1.12
28.5
E
±.006
±.152
1.540*
39.12*
1.540*
39.12*
1.540*
39.12*
1.540*
39.12*
F
±.006
±.152
CFS, CHS
1.23
31.24
1.74
CE, CBE
44.2
1.23
CFS, CHS
31.24
1.74
CFE, CHE
44.2
1.417
36.0
1.417
36.0
* with K extender (CA430) add to height .13
3.3
Panel Cutout
CE, CBE
†
Panel Cutout for thickness of .031-.098 (.8-2.5).
Not recommended for plastic panels.
For snap-in applications, the D sides of the cutout must have a
.02(.508) radius on the installation side.
Snap-in models allow front mounting only.
Ordering Information
Consult your local Corcom sales representative for pricing.
10 C K F S 1
Mounting holes:
.13
Dia. (2) with .23 Dia. x 90° Counter sunk
3.3
Current Rating
(1, 3, 6, 10A)
Series
Terminal Type
(1/4” quick connect)
Mounting Type
S - Snap-in
E - Flange mount
Filter Type
Blank - Unfiltered
B - Pre-wired switch
F - Standard Filter
H - Medical grade filter
CFE, CHE
Extender Option
Blank - Without extender
K - Pre-assembled with
CA430 extender
Available Part Numbers
10CS1
10CKS1
1CFS1
1CKFS1
3CFS1
3CKFS1
6CFS1
6CKFS1
10CFS1
10CKFS1
10CBS1
10CKBS1
10CE1
10CKE1
1CFE1
1CKFE1
3CFE1
3CKFE1
6CFE1
6CKFE1
10CFE1
10CKFE1
10CBE1
10CKBE1
1CHS1
1CKHS1
3CHS1
3CKHS1
6CHS1
6CKHS1
10CHS1
10CKHS1
1CHE1
1CKHE1
3CHE1
3CKHE1
6CHE1
6CKHE1
10CHE1
10CKHE1
Mounting holes:
†
.13
Dia. (2) with .23 Dia. x 90° Counter sunk
3.3
C
87
CBS, CBE Pre-wired Terminals
(Side View)
Libertyville, IL (847) 680-7400 - www.cor.com