MF010 Series Low
Temperature and
MF020 Series High
Temperature
Oxygen Sensors
DESCRIPTION
The MF010 Series Low Temperature Oxygen Sensors is
designed to determine the oxygen concentration in air or inert
gas mixtures with temperatures of -100 °C to 250 °C [-148 °F
to 482 °F] max. The MF020 Series High Temperature Oxygen
Sensors is designed to determine the oxygen concentration in
air or inert gas mixtures with temperatures of -100 °C to 400 °C
[-148 °F to 752 °F] max.
These products are particularly suitable for measuring oxygen
in areas that are not easily accessible, or in closed systems.
The MF010 and MF020 Series may both be supplied with a
measuring range of 0.1 vol%O
2
to 25 vol%O
2
and 0.1 vol%O
2
to 100 vol%O
2
. The entire measurement range is linear for
both series.
The sensor detects the gas mixture’s oxygen partial pressure
(absolute oxygen content) by using a dynamic process that
occurs on two zirconium dioxide discs which form a
hermetically sealed chamber. Because no chemical
substances are used, the sensor exhibits significantly longer
service life than an electrochemical sensor.
The actual oxygen sensor is mounted in the tip of the bar
probe and is protected by a stainless-steel sintered disk which
serves as a flame back-flash stop. The water-proof plastic
housing accommodates the electronics and is mechanically
connected to the bar probe.
FEATURES
No moving parts
Low power consumption
No need for temperature stabilization
Functional testing and calibration in ambient air
Long life
Sensor electronics include:
1. Signal amplifier
2. Control for the ionic pump with an analog component
3. Test component
4. Internal monitoring logic component
5. Power supply for the sensor heating element and the
analog and digital components
6. Reset
7. Voltage monitor
8. Analog output
9. Bi-directional digital output
10. Internal potentiometer
The sensor outputs the measured values via an analog
channel (4 mA to 20 mA or 0 V to 10 V) and a digital channel
(also provides any error signals) which are evaluated and
further processed in a user-specified external mounting device
such as a fuel controller, display, measuring instrument,
programmable logic controller or ventilation system.
Both series may be calibrated manually or self-calibrated using
atmospheric air instead of a reference gas. The sensor can
also conduct external operational self-tests.
Both series may also operate in a fault-proof mode. The sensor
self-monitors during operation and provides error signal output.
A second oxygen sensor is not necessary.
POTENTIAL APPLICATIONS
Boiler combustion controls
Aircraft oxygen generation systems
MF010 Series and MF020 Series
Table 1. MF010 Series Specifications and Order Information
Characteristic
Catalog Listing
MF010-0-LC1
MF010-0-LC2
MF010-0-LC3
MF010-0-LC4
MF010-1-LC1
MF010-1-LC2
MF010-1-LC3
MF010-1-LC4
MF010-2-LC1
MF010-2-LC2
MF010-2-LC3
MF010-2-LC4
Temperature range at probe tip
-100 °C to 250 °C [-148 °F to 482 °F]
Output
0 Vdc to 10 Vdc
4 mA to 20 mA
O
2
range (O
2
partial pressure)
1 mbar to 250 mbar
1 mbar to 1000 mbar
1 mbar to 250 mbar
1 mbar to 1000 mbar
with equivalent O
2
output
0.1 vol%O
2
to 25 vol%O
2
0.1 vol%O
2
to 100 vol%O
2
0.1 vol%O
2
to 25 vol%O
2
0.1 vol%O
2
to 100 vol%O
2
Probe length:
-0-LXX
220 mm [8.7 in]
-1-LXX
400 mm [15.7 in]
-2-LXX
600 mm [23.6 in]
Table 2. MF020 Series Specifications and Order Information
Characteristic
Catalog Listing
MF020-0-LC1
MF020-0-LC2
MF020-0-LC3
MF020-0-LC4
MF020-1-LC1
MF020-1-LC2
MF020-1-LC3
MF020-1-LC4
MF020-2-LC1
MF020-2-LC2
MF020-2-LC3
MF020-2-LC4
Temperature range at probe tip
-100 °C to 400 °C [-148 °F to 752 °F]
Output
0 Vdc to 10 Vdc
4 mA to 20 mA
O
2
range (O
2
partial pressure)
1 mbar to 250 mbar
1 mbar to 1000 mbar
1 mbar to 250 mbar
1 mbar to 1000 mbar
with equivalent O
2
output
0.1 vol%O
2
to 25 vol%O
2
0.1 vol%O
2
to 100 vol%O
2
0.1 vol%O
2
to 25 vol%O
2
0.1 vol%O
2
to 100 vol%O
2
Probe length:
-0-LXX
220 mm [8.7 in]
-1-LXX
400 mm [15.7 in]
-2-LXX
600 mm [23.6 in]
Table 3. MF010 and MF020 Series General Specifications
Characteristic
Supply voltage
Supply current
Ambient temperature (includes solar radiation
Resolution
Heat-up time
Accuracy
Reproducibility
Gas flow rate
Reaction time
Incidental permissible acceleration
Repetitive permissible acceleration
Bar probe material
Housing material
Housing sealing
Connector sealing
Weight (without bar probe)
Termination (female connector)
Certification
Parameter
24 Vdc ±5%
500 mA
-10° C to 50° C [14 °F to 122 °F]
12 bit
approx. 10 min
±2% FS (Full Scale)
±1% FS (Full Scale)
0 m/s to 10 m/s
approx. 3 s
30 g
5g
stainless steel 1.4301
Makrolon 8030 (30% GV), UL94 V-1
IP65
IP67
150 g [5.3 oz]
99-4226-14-07 (Binder 693 series)
complies with EMC directives EN 50082-2 and EN 50081-1 and directives
89/336/EWG and 92/31/EWG
2
www.honeywell.com/sensing
Oxygen Sensors
Figure 1. Dimensional Drawing and Pinout (For reference only: mm/in.)
Varies with catalog listing.
See Table 1.)
88,0
[3.46]
21,0
[0.83]
105,0
[4.13]
Pin
27,7
[1.09]
17,5
[0.69]
30,0
[1.18] 35,5
[1.38] 42,5
[1.67]
1
2
3
4
5
6
+24 Vdc
0V
not assigned
input +24 Vdc
Channel 1
(4 mA to 20 mA)
Channel 2 (I/O)
calibration switch
6
5
4
7
1
2
3
0 V or 4 mA = 0 Vol%O
2
10 V or 20 mA = 25 Vol%O
2
5V
0V
60,0
[2.36]
14,0
[0.55]
7
ground
Ø26,5
[1.04]
Ø12,0
[0.47]
Ø9,0
[0.35]
21,0
[0.83]
42,0
[1.65]
Ø0,21
[5.3]
Figure 2. System Block Diagram
Figure 3. Sensor Mounting
Partition
Atmospheric Air
50 °C [122 °F] Max.
15°
Bar Probe
Gas Mixture
250 °C [482 °F] Max. (MF010)
400 °C [752 °F] Max. (MF020)
Housing
MF010/MF020
Sensor
Ground
Test
1 kOhm + LED
0V
User-Specified
External Monitoring
Device
(4 mA to 20 mA
or 0 V to 10 V
Output)
Power
Supply
24 Vdc
Partition
Calibration
Connection Cable
Bar Probe
O-Ring
Nut
Weld Socket
Order Guide
Catalog Listing
MF010 Series
MF020 Series
Description
Low temperature oxygen sensor (See Table 1 for specifics.)
High temperature oxygen sensor (See Table 2 for specifics.)
Honeywell Sensing and Control
3
WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency stop
devices or in any other application where failure of the
product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
The information presented in this product sheet is for
reference only. Do not use this document as a product
installation guide.
Complete installation, operation, and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of
defective materials and faulty workmanship. Honeywell’s
standard product warranty applies unless agreed to otherwise
by Honeywell in writing; please refer to your order
acknowledgement or consult your local sales office for specific
warranty details. If warranted goods are returned to Honeywell
during the period of coverage, Honeywell will repair or replace,
at its option, without charge those items it finds defective.
The
foregoing is buyer’s sole remedy and is in lieu of all other
warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special,
or indirect damages.
While we provide application assistance personally, through
our literature and the Honeywell web site, it is up to the
customer to determine the suitability of the product in the
application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this
printing. However, we assume no responsibility for its use.
SALES AND SERVICE
Honeywell serves its customers through a worldwide network
of sales offices, representatives and distributors. For
application assistance, current specifications, pricing or name
of the nearest Authorized Distributor, contact your local sales
office or:
E-mail:
info.sc@honeywell.com
Internet:
www.honeywell.com/sensing
Phone and Fax:
Asia Pacific
Europe
+65 6355-2828
+65 6445-3033 Fax
+44 (0) 1698 481481
+44 (0) 1698 481676 Fax
Latin America +1-305-805-8188
+1-305-883-8257 Fax
USA/Canada +1-800-537-6945
+1-815-235-6847
+1-815-235-6545 Fax
Sensing and Control
Honeywell
1985 Douglas Drive North
Minneapolis, MN 55422
www.honeywell.com/sensing
000670-3-EN IL50 GLO Printed in USA
July 2011
Copyright © 2011 Honeywell International Inc. All rights reserved.