30-48 watt medical & ite switching power supplies

TRUMPower
DESCRIPTION
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
TMC40 SERIES
The TMC40 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 30-48
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both
EN60601-1
and
EN60950-1
safety
standards
improves
design-in time and reduces end equipment compliance costs.
FEATURES
Medical and ITE approvals
Compact size 2” x4” x1.18”
Single, dual and triple outputs
Full-range input 90-264 VAC
Low earth leakage current
Level B emissions
RoHS compliant
SAFETY STANDARD APPROVALS
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth Leakage current:
90-264 VAC
47-63 Hz
0.9 A (rms) for 100 VAC
0.5 A (rms) for 240 VAC
150 µA max. @ 264 VAC, 63 Hz
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
62±5 KHz
80-88% typical except TMC40-T31-3 and
TMC40-T31-5 at 75% typical
12 msec minimum at 110 VAC
±0.5% maximum at full load
62±5 KHz
25 A @ 115 VAC, or 50 A @ 230 VAC, except
80-88% typical
TMC40-T31-3
and
4000 VAC from input to output
TMC40-T31-5 at 75%
1500 VAC from input to ground
typical
500 VAC from output to ground
MTBF:
EMC Performance
EN55011 / EN55022:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
Class B conducted, class B radiated
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms
400,000 hours at full load at 25 C ambient,
calculated per MIL-HDBK-217F
o
OUTPUT SPECIFICATIONS
Output voltage/current:
Maximum output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
See rating chart
See rating chart
2% peak to peak on 5.0 V model,
1% peak to peak on other models.
Provided on output #1 only; set at
112–132% of its nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /°
C
maximum
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
Inrush current:
Withstand voltage:
at 25 C cold start
o
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
-10 C to +70 C
-40 C to +85 C
5% to 95% non-condensing
o
Derate from 100% to +50 C linearly to
o
50% at +70 C
o
o
o
o
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com
Website: www.TRUMPower.com
You may not be able to imagine the sensors at the Qatar World Cup!
[i=s]This post was last edited by wangerxian on 2022-11-27 14:02[/i]On November 21, Beijing time, the world-renowned 2022 World Cup officially kicked off in Qatar, a Middle Eastern country. 32 teams f...
wangerxian Integrated technical exchanges
EEWorld invites you to disassemble and see what is inside a skipping rope: A simple electronic counting skipping rope device and its structure
Simple electronic counting rope skipping device and its structure1. Rope skippingRope skipping is a common exercise in life. It can consume body heat and speed up the burning of fat in the body, there...
lanhua Power technology
OC gate circuit and cascade analysis of pull-up resistor selection
Figure 1 Figure 2 Description in the book: Question (a little unsure): My analysis: When a low level input is applied, the AND terminal is regarded as two diodes connected in series and one diode when...
zzjzzjzzj Analog electronics

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号