Fluke 8808A
Digital Multimeter
Extended Specifications
Making measurements is
as simple as pushing a
button
The Fluke 8808A 5.5 digit
multimeter has a broad range
of functions, measuring volts,
ohms and amps with a basic
V dc accuracy of 0.01 %. It is
remarkably easy to use, even by
unskilled operators, because it
makes the measurements you
perform most often extremely
easy and fast to do.
Six setup buttons on the
8808A front panel operate like
a car radio’s station presets.
Simply set up the meter for a
common measurement, then
Features at a glance
press shift followed by a setup
button (S1 to S6) to save the
•
5.5 digit resolution
setup. Now each time you
•
Basic V dc accuracy of 0.01 %
perform that measurement, you
•
Dual display
simply press the appropriate
•
Dedicated dc leakage current measurement
setup key. It’s that easy!
The setup buttons eliminate
•
2x4 ohms 4-wire measurement technique
the need to follow complex work
•
Six dedicated buttons for fast access to instrument setups
instruction sheets. Operators no
•
Hi.Lo limit compare for Pass/Fail testing
longer need to press multiple
buttons to set up a measurement
•
Fluke 45 remote command emulation
function and range, test limits, or
enter other parameters to make
a measurement.
Whether you are performing
The 8808A display has
built-in enunciators that clearly
functional tests or making
Eliminate production
show the operator whether a
critical measurements on test
mistakes
test passes or fails. The pass/fail
points, using the limit compare
indicators take the guesswork
The Fluke 8808A 5.5 digit
mode with pass/fail indicators
out of testing: the result is either
multimeter dependably performs eliminates production mistakes,
within limits or it’s out.
the most common measurements especially those where results
required by today’s applications. are “on the edge.”
General Specifications
Voltage
100V Setting ..............................................................90 V to 110 V
120V Setting .............................................................108 V to 132 V
220V Setting .............................................................198 V to 242 V
240V Setting .............................................................216 V to 264 V
Frequency ...................................................................47 Hz to 440 Hz
Power Consumption ..................................................15 VA peak (10 W average)
Dimensions
Height..........................................................................88 mm (3.46 in)
Width...........................................................................217 mm (8.56 in)
Depth ..........................................................................297 mm (11.7 in)
Weight.........................................................................2.1 kg (4.6 Ib)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 C to 50 C
Storage....................................................................-40 C to 70 C
Warm Up.................................................................½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10 C)
<90 % (10 C to 28 C)
<75 % (28 C to 40 C)
<45 % (40 C to 50 C)
Storage ...................................................................-40 C to 70 C <95 %
Altitude
Operating ...............................................................2,000 Meters
Storage ........................................................................12,000 Meters
Vibration .....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ............................................................400 ms
External Trigger Delay .............................................<2 ms
External Trigger Jitter ..............................................<1 ms
Trigger Input .............................................................TTL Levels
Trigger Output............................................................5 V max
Math Functions
Min/max, relative, hold, compare and dB functions
Electrical
Input Protection ........................................................1000 V all ranges
Overrange ...................................................................10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
2
Fluke Corporation
Fluke 8808A Digital Multimeter Extended Specifications
Fluke 8808A Digital Multimeter Extended Specifications
2 Fluke Corporation
Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ..........................................................1000 V on any range
Common Mode Rejection..........................................120 dB at 50 or 60 Hz 0.1
Normal Mode Rejection.............................................80 dB at Slow Rate
A/D Nonlinearity ........................................................15 ppm of range
(1 k
unbalance)
Input Bias Current .....................................................<30 pA at 25 C
Settling Considerations.............................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
Input Characteristics
Range
200 mV
2V
20 V
200 V
1000 V
Notes:
[1]
At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M .
Full-Scale
(5-1/2 Digits)
199.999 mV
1.99999 V
19.9999 V
199.999 V
1000.00 V
Resolution
Slow
1 V
10 V
100 V
1 mV
10 mV
Medium
10 V
100 V
1000 V
10 mV
100 mV
Fast
10 V
100 V
1000 V
10 mV
100 mV
Input Impedance
>10 G
>10 G
10 M
10 M
10 M
[1]
[1]
1%
1%
1%
Uncertainty
Range
90 days
23 °C
200 mV
2V
20 V
200 V
1000 V
Notes:
[1]
Uncertainty given as
(% of reading + % of range)
[1]
1 year
23 °C
5°C
0.015 + 0.004
0.015 + 0.003
0.015 + 0.004
0.015 + 0.003
0.015 + 0.003
Temperature Coefficient/°C
Outside 18 – 28 °C
0.0015 + 0.0005
0.001 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
5 °C
0.01 + 0.003
0.01 + 0.002
0.01 + 0.003
0.01 + 0.002
0.01 + 0.002
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Fluke 8808A Digital Multimeter Extended Specifications
Fluke Corporation 3
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input ..........................................................750 V rms or 1000 V peak or 8 x 10 Volts-Hertz product
Measurement Method ...............................................AC-coupled true-rms. Measures the ac component of input with up to
1000 V dc bias on any range.
AC Filter Bandwidth..................................................20 Hz – 100 kHz
Common Mode Rejection..........................................60 dB at 50 Hz or 60 Hz (1 k unbalance)
Maximum Crest Factor .............................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Only applies for non-sinusoid signals
7
Input Characteristics
Full-Scale
(5-1/2 Digits)
199.999 mV
1.99999 V
19.9999 V
199.999 V
750.00 V
Resolution
Input Impedance
Slow
1 V
10 V
100 V
1 mV
10 mV
Medium
10 V
100 V
1000 V
10 mV
100 mV
Uncertainty
Range
200 mV
Frequency
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
20 Hz – 45 Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
90 days
23 °C 5 °C
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
[1]
Range
Fast
10 V
100 V
1000 V
10 mV
100 mV
Temperature
Coefficient/°C
Outside 18 – 28 °C
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
1M
2 % shunted by
<100 pf
200 mV
2V
20 V
200 V
750 V
2V
20 V
200 V
750 V
1 year
23 °C 5 °C
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
Notes:
[1]
Uncertainty given as
(% of reading + % of range)
4
Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
4 Fluke Corporation
Fluke 8808A Digital Multimeter Extended Specifications
Resistance
Specifications are for 4-wire resistance function, or 2-wire resistance with REL. If REL is not used, add 0.2
resistance plus lead resistance.
Measurement Method ...............................................Current source referenced to LO input
Max Lead Resistance (4-wire ohms) ......................10 % of range per lead for 200 , 2 k
other ranges.
Input Protection .........................................................1000 V on all ranges
Input Characteristics
Full-Scale
(5-1/2 Digits)
199.999
1.99999 k
19.9999 k
199.999 k
1.99999 M
19.9999 M
100.000 M
Resolution
Current Source
Slow
0.001
0.01
0.1
1
10
100
1k
Medium
0.01
0.1
1
10
100
1k
10 k
Fast
0.01
0.1
1
10
100
1k
10 k
0.8 mA
0.8 mA
0.08 mA
0.008 mA
0.9 A
0.16 A
0.16 A || 10 M
ranges. 1 k
for 2-wire
per lead on all
Range
200
2k
20 k
200 k
2M
20 M
100 M
Uncertainty
Range
90 days
23 °C
200
2k
20 k
200 k
2M
20 M
100 M
Notes:
[1]
Uncertainty given as
(% of reading + % of range)
[1]
1 year
23 °C
5 °C
0.03 + 0.004
0.02 + 0.003
0.02 + 0.003
0.02 + 0.003
0.04 + 0.004
0.25 + 0.003
1.75 + 0.004
Temperature Coefficient/°C
Outside 18 – 28 °C
0.003 + 0.0006
0.003 + 0.0005
0.003 + 0.0005
0.003 + 0.0005
0.004 + 0.0005
0.01 + 0.0005
0.2 + 0.0005
5 °C
0.02 + 0.004
0.015 + 0.002
0.015 + 0.002
0.015 + 0.002
0.03 + 0.003
0.2 + 0.003
1.5 + 0.004
DC Current
Input Protection ........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses.
Shunt Resistance .......................................................0.01 for 2 A and 10 A ranges
1 for 20 mA and 200 mA
Burden voltage < 5 mV for 200 A and 2 mA range.
Input Characteristics
Full-Scale
(5-1/2 Digits)
199.999 A
1999.99 A
19.9999 mA
199.999 mA
1.99999 A
10.0000 A
Resolution
Slow
0.001 A
0.01 A
0.1 A
1 A
10 A
100 A
Medium
0.01 A
0.1 A
1 A
10 A
100 A
1 mA
Fast
0.01 A
0.1 A
1 A
10 A
100 A
1 mA
<5 mV
<5 mV
<0.05 V
<0.5 V
<0.1 V
<0.5 V
Burden Voltage
Range
200 A
2 mA
20 mA
200 mA
2A
10 A
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 5
Fluke 8808A Digital Multimeter Extended Specifications
Fluke Corporation 5