Corcom Product Guide
Versatile Power Entry Module with Small Footprint
Catalog: 1654001
Issue Date: 06.2011
P Series
UL Recognized
CSA Certified
VDE Approved
PE filtered
PS non-filtered
PS filtered
P Series
The P series CHAMELEON power entry module offers
the most popular features in a small footprint design
As the first 10A module to provide all five power
entry functions in one compact design, the
chameleon module readily adapts to its environment
and the needs of international markets.
• Snap-in or flange mounting
• Standard IEC 60321-1 C14 power inlet
• Both North American and metric fusing
capabilities
• Two voltage selection options (for 4-voltage
selection, see the M, L or LA Series)
• Optional DPST on/off switch
• Filter options for general purpose, medical and
high-performance EMI filtering
The CHAMELEON module’s compact design and
modular construction allows selection of the required
power entry feature — without altering the panel
cutout. And the CHAMELEON module, with its
optional adapters, will fit several common panel
cutouts.
L models feature a high performance medical filter
designed to help bring most digital equipment
(including switching power supplies) into compliance
with EN55022, Level B (as well as FCC part 15J, Class
B) conducted emissions limits. They are available
with current ratings of 6 and 10A. These high
performance versions are only available with
mounting ears, single voltage selection, in a complete
RFI shield with options for switch, fuses and current
ratings. Mounting extenders are not compatible with
the L or Z models.
Z models provide a high performance three element
differential mode circuit filter, with extended EMI
attenuation similar to the M Series Z models, to help
bring most digital equipment (including switching
power supplies) into compliance with EN55022, Level
B (as well as FCC Part 15J, Class B) conducted emis-
sions limits. They are available with current ratings of
6 and 10A. These high performance versions are only
available with mounting ears, single voltage selection,
in a complete RFI shield with options for switch, fuses
and current ratings. Mounting extenders are not
compatible with the L or Z models. For minimum
depth behind the panel, see the M Series
B models are non-filtered and incorporate an inter-
connection block. The block connects the voltage
selection terminals of an unfiltered CHAMELEON
module with an IEC connector and an optional switch
to reduce external wiring. Compatible with the A or B
RFI shield options.
Filter Types
The CHAMELEON module has four filter and one
non-filtered option:
S models provide an extended performance two
element circuit EMI filter, with attenuation similar to
the EEB Series power inlet filter. It offers protection
for general purpose applications with stray Line to
Ground and Line to Line noise that must be attenu-
ated at the power inlet. These filters have limited
leakage current and are available in current ratings of
3, 6 and 10A.
H models provide susceptibility protection with
minimal leakage current, and are suitable for patient
care and non-patient care medical equipment.
192
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
Corcom Product Guide
Versatile Power Entry Module with Small Footprint
Catalog: 1654001
Issue Date: 06.2011
(continued)
P Series
Ordering Information
Part numbers are constructed by selecting the alphanumeric character which represents the desired feature.
Note: For any option where shown as “0” use the digit ZERO (0) not the letter (O).
Mounting
Style
Extender
Options
Switch
Options*
Voltage
Select
Fuse
Options
Filter
Type
Filter
Current
Shield
Options
P S J 0 X S S 6 0
Mounting Style
E = Mounting ears
(Horizontal axis)
S = Snap-in
M = Mounting ears
(Vertical axis)
Switch Option
0 = Blank
S = Double pole
single throw
(DPST)
Fuse Options
D = Dual fuse
S = Single fuse
Filter Current
0 = No filter
3 = 3A
6 = 6A
X = 10A
3
Power Inlet Filters & Power Entry Modules
Extender
Option
0 = None
C = FN260
J = J Series
L = L Series
Input Voltage Select
0 = Single voltage
S = Dual voltage
select (SMPS)
X = Dual voltage
select
(series/parallel)
Shield Options
Filter Type
0 = No shield
0 = No filter
A = Filter shield
B = Interconnection
B = Complete
block
Shield
H = Medical Filter
L = High Performance C = Complete can
(for L & Z models)
Medical
4
S = General purpose
Z = High Performance
4
Extender for
J-Series or others
Fuse holder
Notes:
1
L Series extender cannot be added to units
with a shield. No style of extender can be
added to units with B or C shields.
Dual voltage options are not available with
L or Z Filter Types
2
3 When using the interconnection block, the last
3 digits of the part number are BX (0, A, or B)
Optional switch
4 High performance versions (L or Z filter types)
are available with any switch or fuse option but
only in 6 or 10A with horizontal (PE) or
vertical (PM) mounting ears, single voltage (0),
complete shield (C) and no extenders
*
For alternative switch orientation options,
please contact technical support or your
Corcom product sales representative
Optional filter
The part number PS0SXSS6B would represent:
P Series (P) with a snap-in mount (S) with no extender (0) a switch (S) dual voltage select (X) single fusing (S)
general purpose filter (S) for 6A (6) with a B shield (B)
193
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
Corcom Product Guide
Versatile Power Entry Module with Small Footprint
Catalog: 1654001
Issue Date: 06.2011
(continued)
P Series
Voltage Selection
P series power entry modules include the voltage
selector integral with the fuse holder. Three voltage
selection options are each supported by one of three
different fuse holders. The fifth digit of the part
number specifies which of the three fuse holders is
included to provide the desired voltage selection.
The single voltage fuse holder (option “0”) has no
voltage indication markings. The dual voltage options
select 115V or 230V by removing the fuse holder, flip-
ping it over, and reinstalling it. Voltage selection is
indicated through a window in the P Series door. The
“SMPS” fuse holder (option “S”) jumpers two inde-
pendent P Series terminals to indicate 230V
operation to a switching mode power supply. The
“PRSR” parallel/serial fuse holder (option “X”)
connects the windings of the equipment’s dual
primary transformer (not included) to step down the
voltage or double up the current. The markings on
the voltage selection fuse holders also remind the
user to install the appropriate fuse for the current at
the selected voltage.
Shield Options
The P series offers several RF shield options. The
metal shield, optional on S, H and B filtered models,
provides shielding from radiated emissions and
provides an RF ground for the filter to the panel. This
shield is available in two versions; a shield of the filter
components (designated by an A as the final digit)
and a complete shield (designated by B as the final
digit).
The A shield covers the filter portion of the module
and increases performance of the filter by protecting
the components from RFI coupling. This shield
allows the use of the C or J extender.
The B shield covers the entire power entry module
with metal, protecting the filter from RFI coupling,
and covering the mounting cut-out to block RFI
entering or leaving the equipment. The B shield
cannot be used with any extender.
A complete metal enclosure is integral to both the
high performance L and Z models, and must be spec-
ified by a C in the part number’s final digit. This
option is only available with the L or Z models.
Input Voltage Selection Schemes
S - “SPMS” Jumper Type
115V
Power on/off
Hot
Note 1
Ground
Note 1
Neutral
Note 1
H
C
D
G
A
B
230V
Power
input
Sel.
Jumpers
SMPS
Note 1
Provision for
single fusing
X - “PRSR” Parallel / Serial Type for
Dual Primary Transformer
115V
Power on/off
Hot
Note 1
Ground
Note 1
Neutral
Note 2
H
C
D
G
A
B
230V
“A” Shield
115V
115V
Transformer
primary
Note 2
“B” Shield
Provision for
single fusing
½½
½
194
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
Corcom Product Guide
Versatile Power Entry Module with Small Footprint
Catalog: 1654001
Issue Date: 06.2011
(continued)
P Series
Fuseholder
Another feature of the P series power entry module is
the versatile fusing arrangement. The fuse holder can
hold two 1/4” x 1-1/4” (3AG) or 5 x 20mm (metric)
fuses. Single fusing is supported with a conversion
clip that shorts one of the two fuse positions, and is
designated by an S in the sixth part number digit. A
module designated for a single fuse may be reconfig-
ured by the manufacturer or the user to accept two
fuses by simply removing the shorting clip. For appli-
cations intended for dual fusing, specify a D in the
sixth part number digit.
Conversion Clip for
Single Fuse
Applications
Accessories
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord
PA100: Power interconnect assembly
For voltage select models. Designed for use with
either filtered or non-filtered units, 6” wire leads
3
Power Inlet Filters & Power Entry Modules
PA101: Plug only
PA102: Pins only for use with PA101
PA105: Same as PA100 but with two wires for
units with no voltage selection
Interconnection Block
Installation of the unfiltered versions of the P series
requires wiring of the IEC socket terminals to the
optional switch and the switch to the fuse holder.
Labor can be eliminated by ordering the module
with an interconnection block. This feature, desig-
nated by “BX” in the seventh and eighth digits, pre
wires the module so that only connection to the
equipment must be done during installation. The
interconnection block includes a plastic case to
prevent access to the internal connections.
The dimensions of this alternative are the same as the
filtered versions.
PA400: J Extender
Extends P Series height to fit J panel cutout
PA410: L Extender
Extends P Series width to fit L panel cutout
PA420: C Extender
Extends P Series height to fit C panel cutout
PA400
Extender
Extender
J Series
for
J-Series or others
Fuse holder
C & L Extenders can not be used with B Shields.
L Extender can not be used with shields
Optional switch
195
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
Optional filter
Corcom Product Guide
Versatile Power Entry Module with Small Footprint
Catalog: 1654001
Issue Date: 06.2011
(continued)
P Series
Specifications
Maximum leakage current each Line to Ground:
H & L Models
S & Z Models
@ 120 VAC 60 Hz:
2
A
.25 mA
@ 250 VAC 50 Hz:
5
A
.50 mA
Hipot rating (one minute):
Line to Ground:
Line to Line:
Rated Voltage(max.):
Operating Voltages:
Selectable or Fixed
Operating Frequency:
Rated Current:
2250 VDC
1450 VDC
250VAC
115/230 VAC
50/60 Hz
Non-Filtered – 10A
Filtered – 3, 6 or 10A
Electrical Schematics
H Model
LINE
LOAD
S Model
LINE
LOAD
L Model
Fuseholder:
Accepts one or two fuses
.25 x 1.25”(not included)
or 5 x 20mm (not included)
DPST
10,000 operations at 51A max. inrush
LINE
LOAD
Switch:
Z Model
LINE
LOAD
Case Styles
Front View
PM
High Performance
Version Only
PE
PS
2.496
63.4
2.88
2.39
73.15
60.71
2.29
58.16
½
1.22
30.98
1.58
40.13
1.95
49.53
1.22
30.98
196
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com