DC/DC Converters
THI 2M Series, 2 Watt
Features
◆
Ultracompact DIP 16 package
◆
I/O isolation 3000 VACrms rated for
300 Vrms working voltage
CB
IEC 60950-1
IEC 60601-1
Scheme
UL 60950-1
UL 60601-1
◆
Medical safety to UL 60601-1 and
IEC/EN 60601-1 3rd edition, 2 x MOOP
◆
Industrial safety to IEC/EN/UL 60950-1
◆
Operating temp. range –40°C to +71°C
◆
3-years product warranty
The THI 2M series is a new range of ultra-compact 2W DC/DC-converters providing
a high I/O-isolation voltage of 3000 VAC. With a reinforced I/O-isolation system
this product is an economical solution for many applications in instrumentation,
industrial controls, medical equipment and everywhere where supplementary- or
reinforced insulation is required to meet requested safety standards.
Full SMD-design with exclusive use of ceramic capacitors ensure a very high reli-
ability and a long product lifetime.
Models
Order code
THI 2-0511M
THI 2-0512M
THI 2-0513M
THI 2-0522M
THI 2-0523M
THI 2-1211M
THI 2-1212M
THI 2-1213M
THI 2-1222M
THI 2-1223M
THI 2-2411M
THI 2-2412M
THI 2-2413M
THI 2-2422M
THI 2-2423M
24 VDC ± 10%
(nominal 24 VDC)
12.0 VDC ± 10%
(nominal 12 VDC)
5.0 VDC ± 10%
(nominal 5 VDC)
Input voltage range
Output voltage
5 VDC
12 VDC
15 VDC
±12 VDC
±15 VDC
5 VDC
12 VDC
15 VDC
±12 VDC
±15 VDC
5 VDC
12 VDC
15 VDC
±12 VDC
±15 VDC
Output current max.
400 mA
165 mA
133 mA
±83 mA
±66 mA
400 mA
165 mA
133 mA
±83 mA
±66 mA
400 mA
165 mA
133 mA
±83 mA
±66 mA
Efficiency typ.
66 %
66 %
66 %
72 %
73 %
66 %
66 %
66 %
74 %
75 %
66 %
66 %
66 %
74 %
75 %
http://www.tracopower.com
Page 1 of 3
DC/DC Converters
THI 2M Series
Input Specifications
Input current no load / full load
5 Vin models:
60 mA / 600 mA typ.
12 Vin models:
30 mA / 250 mA typ.
24 Vin models:
15 mA / 135 mA typ.
0.3 A max.
5 Vin models:
1.0 A
12 Vin models:
0.5 A
24 Vin models:
0.2 A
5 Vin models:
9 V max.
12 Vin models:
18 V max.
24 Vin models:
30 V max.
internal capacitors
±4 %
1 % max.
1.2 % / 1 % change of Vin
10 % max.
12 % max. for 5 Vout models.
150 mVpk-pk max
±0.02 %/K
0.5 sec. max.
2 % of rated max. current
single output models:
330 µF max.
dual output models:
100 µF max.
(each output)
– Operating
– Storage
– Casing
–40°C to +71°C
–40°C to +125°C
+90°C max.
2.5 %/K above 60°C
95 % rel. H max.
>2.0 Mio h
– Input variation
– Load variation 20 – 100 %
2 Watt
Reverse voltage protection
Recommended external input fuse
(slow blow)
Surge voltage
(1 sec. max.)
Input filter
Output Specifications
Voltage set accuracy
Voltage balance
(dual output models)
Regulation
Ripple and noise
(20 MHz Bandwidth)
Temperature coefficient
Short circuit protection
Minimum load
Capacitive load
General Specifications
Temperature ranges
Derating
Humidity
(non condensing)
Reliability, calculated MTBF
(MIL-HDBK-217F, @ 25°C, ground benign)
Isolation voltage
– Input/Output
Reinforced, rated for 300 Vrms working voltage
(50Hz, 60sec)
Medical applications in accordance to IEC/EN 60601-1:
3000 VAC, 2 x MOOP
IT applications in accordance to IEC/EN 60950-1:
4000 VAC
Isolation test voltage
(1 sec.)
Leakage current
(at 240VAC, 60Hz)
Isolation capacity
Isolation resistance
Switching frequency
Safety standards
– Input/Output
– Input/Output
6‘000 Vpk
2 µA max.
20 pF max.
(at 100KHz, 1V)
>10 Gohm
(at 500VDC)
50 – 100 kHz (PFM)
IEC/EN 60950-1, UL 60950-1
CSA C22.2 No. 60950-1-03
IEC/EN 60601-1 3rd edition, 2 x MOOP,
UL 60601-1, CSA C22.2 No. 601.1
– CSA certificate
for medical electrical
for information technology
– CB test certificate
for medical electrical
for information technology
equipment
equipment
equipment
equipment
www.tracopower.com/products/thi2m-csa60601.pdf
www
.
tracopower
.
com
/
products/thi2m-csa60950.pdf
www.tracopower.com/products/thi2m-cb60601.pdf
www.tracopower.com/products/thi2m-cb60950.pdf
Safety approvals
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
http://www.tracopower.com
Page 2 of 3
DC/DC Converters
THI 2M Series
General Specifications
Casing material
Weight
Soldering temperature
non conductive plastic
(UL 94V-0 rated)
5.1 g (0.18oz)
max. 265°C / 10 sec
2 Watt
Application note:
www.tracopower.com/products/thi2m-application.pdf
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.
Outline Dimensions
16
1
Pin-Out
Pin
1
7
8
9
Single
–Vin (GND)
No con.
No con.
+Vout
–Vout
+Vin
Dual
–Vin (GND)
No con.
Common
+Vout
–Vout
+Vin
23.8
bottom view
(0.60)
15.24
(0.94)
2.54
(0.10)
10
9
0.5
(0.02)
3.2
(0.13)
7
8
10
16
(0.12)
3.0
1.77
(0.07)
10.16
13.7
(0.54)
8.6
(0.34)
Dimensions in [mm], () = Inch
Pin diameter: 0.5 ±0.05 (0.024 ±0.002)
Tolerances: ±0.25 (±0.01)
Pin pitch tolerances: ±0.05 (±0.002)
Specifications can be changed without notice! Make sure you are using the latest documentation, downloadable at www.tracopower.com
www.tracopower.com
Rev. July 8. 2013
Page 3 of 3