DATASHEET
Rev. B
PSMP05 SERIES
85~264VAC (125~373VDC) Input Voltage Range
Single Outputs, Isolation Class II
5 Watts, Open Frame
Medical AC/DC Switching Power Supplies
MEDICAL
FEATURES
•
Isolation Class II
•
2.28” x 1.77” x 0.75” Open Frame Package
•
Single Outputs
•
100% Full Load Burn-in Tested
•
Cooling by Free Air Convection
•
±10% Voltage Adjustment
•
RoHS Compliant
•
Energy Star Compliant
•
Green Design, No-Load Power Consumption < 0.3W
•
Withstand 2G Vibration Test
•
Universal Input Voltage: 85~264VAC or 125~373VDC
•
5 Watts Output Power
•
All Using 105°C Long Life Electrolytic Capacitors
•
-20°C ~ +70°C Wide Operating Temperature Range
•
Short Circuit, Over Load, Over Voltage, and Brown-out
(Low AC Input Voltage) Protection
•
UL60601-1, TUV EN60601-1, and IEC60601-1 Medical
Approvals
DESCRIPTION
The PSMP05 series of Medical AC/DC switching mode power supplies provides 5 Watts of continuous output power in a
2.28” x 1.77” x 0.75” open frame package. This series consists of 5V, 12V, 15V, and 24VDC output models with a
universal input voltage range of 85~264VAC or 125~373VDC. These power supplies are protected against short circuit,
over load, over voltage, and brown-out (low AC input voltage) conditions and have an MTBF of 217,400 hours using MIL-
HDBK-217F. This series also has UL60601-1, TUV EN60601-1, and IEC60601-1 medical approvals. All models have been
100% full load burn-in tested and are RoHS and Energy Star compliant.
02/05/2013
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 1 of 4
Rev. B
PSMP05 Series
5 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
SPECIFICATIONS: PSMP05 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
Low Line
High Line
Low Line
Inrush Current
High Line
No Load Power Consumption
OUTPUT SPECIFICATIONS
Output Voltage
Input Current
Voltage Tolerance
Voltage Adjustment Range
Load Regulation
Line Regulation
Output Power
Output Current
Ripple & Noise
(See Note 1)
Low Line
Hold-Up Time
High Line
Setup Time
(See Note 3)
Rise Time
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Load Protection
Short Circuit
Brown-out (Low AC Input Voltage)
GENERAL SPECIFICATIONS
Efficiency
Withstand Voltage
Isolation Resistance
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Vibration
Cooling
MTBF
PHYSICAL SPECIFICATIONS
Weight
Dimensions (L x W x H)
SAFETY & EMC
Safety Approvals
EMI Conduction & Radiation
Harmonic Current
EMS Immunity
5VDC output model
12V, 15V, & 24VDC output models
Low Line to High Line at rated load
0% to 100% rated load
Full Load, Vin = 115VAC
Full Load, Vin = 230VAC
Cold Start, Vin = 115VAC
Cold Start, Vin = 230VAC
AC Input Voltage Range
DC Input Voltage Range
TEST CONDITIONS
Min
85
125
47
0.13
0.08
25
45
0.3
See Table
+2
+1
+10
+1
+0.5
+0.5
5
See Table
See Table
25
100
100
25
+0.03
145
yes
yes
Vin = 230VAC
Input to output
500VDC
With derating (see derating curve)
Non-condensing
See Table
4000
100
VAC
MΩ
Nom
Max
264
373
63
Unit
VAC
VDC
Hz
A
A
W
5VDC output model
12V, 15V, & 24VDC output models
-2
-1
-10
-1
-0.5
-0.5
0
%
%
%
%
W
Full Load, Vin = 115VAC
Full Load, Vin = 230VAC
Full Load, Vin = 115/230VAC
Full Load, Vin = 115/230VAC
0~50°C
Latch-off mode
Hiccup mode, recovers automatically after fault condition is removed
ms
ms
ms
%/°C
%
%
-0.03
115
110
MIL-HDBK-217F
-20
+70
°C
-40
+85
°C
20
90
% RH
10
95
% RH
10~500Hz, 2G 10min/1cycle, period for 60 min each along X, Y, and Z axes
Free air convection
217,400 hours
Approximately 1.16oz (33g)
2.28 x 1.77 x 0.75 inches
(57.9 x 45 x 19.1 mm)
UL60601-1, TUV EN60601-1, and IEC60601-1
EN55011: 2007+A2: 2007 Class B
EN61000-3-2: 2006 Class A, EN61000-3-3: 1995+A1: 2001+A2: 2005
EN60601-1-2: 2001+A1: 2006, IEC61000-4-2,3,4,5,6,8,11 light industry level, criteria A
02/05/2013
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 2 of 4
Rev. B
PSMP05 Series
5 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
MODEL SELECTION TABLE
Model Number
PSMP-05-05
PSMP-05-12
PSMP-05-15
PSMP-05-24
85 ~ 264 VAC
or
125 ~ 373 VDC
Input Voltage Range
Output Voltage
5 VDC
12 VDC
15 VDC
24 VDC
Output Current
1A
0.42A
0.33A
0.21A
Ripple & Noise
(1)
100mVp-p
150mVp-p
150mVp-p
240mVp-p
Efficiency
71%
75%
75%
77%
Output Power
5W
5W
5W
5W
NOTES
1. Ripple & noise is measured at 20MHz bandwidth by using 12" twisted pair-wire terminated with 0.1µF and 47µF
capacitors in parallel.
2. Tolerance includes set up tolerance, line regulation, and load regulation.
3. The length of the setup time is measured a first cold start. Turning the power supply ON and OFF very quickly may lead
to an increase in the setup time.
4. The power supply is considered a component which will be installed into final equipment. The final equipment must be
re-confirmed that it still meets EMC directives.
BLOCK DIAGRAM
I/P
EMI
FILTER
RECTIFIERS
&
FILTER
POWER
SWITCHING
PWM
CONTROL
RECTIFIERS
&
FILTER
DETECTION
CIRCUIT
+V
-V
DERATING CURVE
Load vs Temperature
Load vs Input Voltage
Ta=25°C
100
80
100
90
Load (%)
Load (%)
80
70
60
50
60
40
20
40
85
100
110
120
264
-20
0
10
20
30
40
50
60
70
Ambient Temperature (°C)
Input Voltage (VAC)
02/05/2013
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 3 of 4
Rev. B
PSMP05 Series
5 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
MECHANICAL DRAWING
Unit: inches (mm)
0.75 (19.1)
0.07 (1.8)
0.041 (1.05)
0.177 (4.5)
0.138 (3.5)
AC INPUT
C2
DC OUTPUT
0.689”
(17.5mm)
AC/N
AC/L
0.196 (5)
0.21 (5.33)
1.85 (47)
2.28 (57.9)
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers
will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just
one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with
its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing
support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects.
Contact
Wall Industries
for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
(603)778-2300
(888)597-9255
(603)778-9797
sales@wallindustries.com
www.wallindustries.com
5 Watson Brook Rd.
Exeter, NH 03833
02/05/2013
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
0.491
(12.47)
+V
0.789 (20.04)
-V
1.77 (45)
Page 4 of 4