FEATURES:
ACCOMMODATE UNIVERSAL AC SOURCES
COMPACT 4.5”x 2.5”x1.2” PCB
FORMAT
MEET MEDICAL STANDARD EN 60601-1/
UL 60601-1
EMI MEET EN 55011 & FCC CLASS B
CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage:
90-264Vac.
Input Connector:
V-M connector.
Input Frequency:
47-63Hz.
Inrush Current:
Typ. 9.68Arms @230Vac.
Input Current:
Typ. 1.8A@115Vac, 0.9A @230Vac.
Dielectric Withstand:
Meet IEC 60601-1.
EMI:
Meet EN55011 & CISPR Class B.
Hold-up Time:
Typ. 12.8mS @115Vac, 76mS @ 230Vac.
Earth Leakage:
Less than 0.189mA @230Vac.
OUTPUT SPECIFICATION
Output Voltage:
See Ratings Chart.
Output Current:
See Ratings Chart.
Output Connector:
V-M connector.
Output Wattage:
Typ. 100 watts.
Line Regulation:
Typ. ±0.5%.
Load Regulation:
Main VO1 typ ±1%.
Aux. VO2 typ. ±1%.
Aux. VO3 typ. ±5%.
Aux. VO4 typ. ±5%.
Noise & Ripple:
Typ. 1.0% peak to peak.
OVP:
Built-in at main output (Latch).
Adjustability:
Available at VO1 and VO2.
Overload Protection (OLP):
Fully protected against outputs overload and short circuit.
Typ. OLP set at 110-150% of rating output power.
Consult the factory for OLP setting.
Operating Temperature:
0 to +50 °C under forced air flow
for full load.(Various with output voltage).
Storage Temperature:
-20 to +85°C.
Temperature Coefficient:
Typ. 0.05%/°C.
Cooling:
Recommended 24CFM cooling fan for
full rating load.
Power Density:
7.4 Watts / Cubic Inch.
Construction:
4.5”x2.5“x1.2” PCB format.
GENERAL SPECIFICATION
Efficiency:
Typ. 74-80%(Various with output voltage).
Switching Frequency:
68 KHz.
Circuit Topology:
Single-Forward circuit.
Multiple-Fixed freq. flyback circuit.
Transient Response:
Output voltage returns in less than
14-44mS following a 25% load change.
Safety Standard:
Meet Medical standards UL 60601-1/
EN 60601-1 Class I.