Cable breakage measuring instrument circuit

Source: InternetPublisher:无人共我 Keywords: measuring instrument cable Updated: 2021/02/12

25.<strong>Cable</strong>Breakage<strong>Measuring Instrument</strong><strong>Circuit</strong>.gif

This device uses a single probe to test whether multiple paths are connected or disconnected and indicates which
end is broken, as shown in Figure 5-25.
    Components such as 4060 form a 20kHz oscillator, and the connected 100kfl and
200pF resistor-capacitor components can change this frequency. Each output of the IC, such as Q3 and Q4,
can drive a light-emitting diode or a piezoelectric buzzer to provide necessary indications. This
circuit is installed in a metal box. Hold the cable with one hand and
the tester with the other hand, and touch each pin of the cable with the probe in sequence . If the pin is
connected, stray capacitance will reduce the tone frequency.
Perform the same test on the other end of the cable . The break should be at the end of the frequency that is least affected by stray capacitance
. Touch one pin of the cable with the probe and hold the other end of the cable with your other hand
to check one-on-one. If the cable is intact, the oscillator will stop oscillating. Short circuits in the internal coils of motors and large inductors can also
be checked by comparing them with each other.


EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

About Us Customer Service Contact Information Datasheet Sitemap LatestNews


Room 1530, 15th Floor, Building B, No.18 Zhongguancun Street, Haidian District, Beijing, Postal Code: 100190 China Telephone: 008610 8235 0740

Copyright © 2005-2024 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号